…13 Specification
Configuration security
Hardware switch configuration
Access to configuration is enabled only after a hardware
switch has been set. This switch is situated behind a
tamper-evident seal.
Password protection
Access to configuration levels is enabled only after the
user has entered a password:
• Calibrate level: user-assigned password
• Advanced level: user-assigned password
• Service level: factory set password
Input
pH/ORP/plon sensor types
pH: Glass, Antimony (Sb)
ORP: (Redox): Platinum (Pt), Gold (Au)
pIon: Custom user-programmable
Input impedance
>1x10
13
Ω
pH/ORP/plon measurement range and resolution
Type
Range
Display
resolution
Accuracy
repeatability
pH
0 to 14 pH
(–2 to 16 over range)
0.01 pH
±0.01 pH
ORP
–1500 to 1500 mV
1 mV
±1 mV
pIon
–1500 to 1500 mV
1 mV
±1 mV
Dynamic response
<1 second for 90 % step change at 0 seconds damping
Damping
Configurable: 0 to 99.9 seconds
Conductivity sensor types
AWT210: ABB 2-electrode conductivity sensors
AWT210: ABB 4-electrode conductivity sensors
AWT210: ABB toroidal conductivity sensors
Conductivity measurement range and resolution
AWT210 2-electrode conductivity transmitter:
Cell
constant
Conductivity
range
Display
resolution
Accuracy
repeatability
0.01
0 to 200 µS/cm
0.001 µS/cm
±1.0 % of
measurement
range per
decade
0.1
0 to 2000 µS/cm
0.01 µS/cm
1
0 to 20000 µS/cm
0.1 µS/cm
…Conductivity measurement range and resolution
AWT210 4-electrode conductivity transmitter:
Sensor
group
Conductivity
range
Display
resolution
Accuracy
repeatability
A
0 to 2000 mS/cm
0.1 µS/cm
±0.5 % of
measurement
range per decade
B
0 to 2000 µS/cm
0.01 µS/cm
AWT210 toroidal conductivity transmitter:
Sensor
Conductivity
range
Display
resolution
Accuracy
repeatability
ABB
toroidal
0 to 2000 mS/cm
1.0 µS/cm
±0.5 % of
measurement
range per decade
Temperature input
Temperature element types
Pt100 (2 or 3-wire) Automatic temperature compensation
Pt1000 (2 or 3-wire) Automatic temperature compensation
3k Balco (2 or 3-wire) Automatic temperature compensation
None
Manual temperature compensation
Measurement range and resolution
Temperature
element
Temperature
range
Accuracy
Repeatability
Pt100
–20 to 200 °C
(–4 to 392 °F)
±0.1 °C (±0.18 °F)
after calibration
Pt1000
3K Balco
None
User-programmable
20 to 300 °C (–4 to 572°F)
N/A
pH/ORP/plon temperature compensation modes
Type
Manual
Automatic
Nernstian
Nernstian
with solution
coefficient
Solution
compensation
coefficient
pH
ORP
pIon
Conductivity temperature compensation modes
Temperature
element
AWT210
2-electrode
AWT210
4-electrode
AWT210
toroidal
0 to 15 % NaOH
0 to 20 % NaCI
0 to 18 % HCl
0 to 20 % H2SO4
Pure water neutral salt
Pure water trace base
Pure water trace acid
User-defined