9 Password security and Access Level
Passwords are entered at the
Enter Password
screen accessed
via the
Access Level
– see below.
Access Level
The
Access Level
is entered via the
Operator
/
Enter
Configuration
menu option. Use the
keys to highlight
the required level and press to enter the level.
Figure 20 Access level screen
Level
Access
Logout
Displayed only after
Calibrate
or
Advanced
levels are accessed.
Logs the user out of the current level. If passwords are set, a
password must be entered to access these levels again after
selecting
Logout
.
Read Only
View all parameters in read-only mode.
Calibrate
Enables access and adjustment of
Calibrate
level only
(calibration menus are sensor-specific).
Advanced
Enables configuration access to all parameters.
Service
Reserved for authorized service technicians only.
Table 3 Access level menu details
Figure 21 Enter password screen
Write protect switch
When the
Write Protect
switch (see page 6) is in the
ON
position, the transmitter is write-protected (and the Write
Protected icon is displayed – see page 19). This means
that only the
Read Only
access level is available to the operator.
When this switch is in the
OFF
position, all access levels are
available (
Read Only, Calibrate, Advanced
and
Service
).
Setting passwords
Passwords can be set to enable secure access at 2 levels:
Calibrate
and
Advanced
. The
Service Level
is password
protected at the factory and reserved for factory use only.
Passwords can contain up to 6 characters and are set, changed
or restored to their default settings at the
Device Setup/
Security Setup
parameter – see page
39 .
Note. The transmitter is supplied with blank passwords for the
Calibrate
and
Advanced
levels, therefore, the
Calibrate
and
Advanced
levels levels can be accessed without password
protection. It is recommended to set passwords for these
access levels.
Password recovery
Advanced level password recovery
To recover the
Advanced
level password, move the
Write
Protect
switch to the
OFF
position (see page 6). Select the
Service Access
level and enter the
Service
level password to
gain access. From the
Service
level, the
Device Setup
menu can
be accessed to reset the
Advanced
level password.
Service level password recovery
If the
Service
level password is lost, the only way to recover it is
by following the procedure to reset all parameters to the
factory default values as described in Appendix B, page 52.
This resets all parameters including passwords.
Access Level
Calibrate
Advanced
Service
Read Only
Back
Select
Cursor/Password character indicator (maximum 6 characters)
Cursor – scroll characters using the
keys;
press (Next) to accept character; press (OK) to
accept password while last character is highlighted
Enter Password