INTRODUCTION
SPECIFICATIONS
1 - 9
Table 1-4. Optional Potentiometric Position Transmitter
Specifications (Type AV33
1 Only)
Property
Characteristic/Value
Total Resistance
2,000
Ω ±
20%
Power Rating
1 W up to 70
°
C (158
°
F)
0 W at or above 125
°
C (257
°
F)
Wiper Rate of
Change
9.9
Ω
nominal per degree of cam rotation
Temperature Effect 0.05% (500 ppm) per
°
C
0.03% (278 ppm) per
°
F maximum
Maximum Voltage
35 VDC or 30 VAC across the potentiometer ends
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE
Table 1-5. 4 to 20 mA Position Transmitter Specifications
(Type AV33
2
and AV44
0 )
Property
Characteristic/Value
Supply Voltage
16 to 34 VDC
Output Signal
4 to 20 mA
Output Loading
500
Ω
at 24 VDC
1000
Ω
at 34 VDC
Accuracy
0.6% of span maximum
Hysteresis
0.5% of span maximum
Repeatability
0.4% of span maximum
Ambient Tempera-
ture Effect
0.063% per
°
C
0.035% per
°
F
EMI/RFI Effect
1.5% maximum at 10 volts per meter field
strength, 20 to 450 MHz
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE
Table 1-6. Agency Approvals
Nomenclature
Approval/Certification
AV3/4
1, 2
FM and CSA approval and certifications in the
following categories:
Nonincendive for Class I, Division 2,
Groups A, B, C, D.
Intrinsically safe for Class I, Division 1,
Groups A, B, C, D, when used with an
approved intrinsic safety barrier.
NOTES:
1. Approvals and certifications pending. The position transmitters of Type AV3/4
1/2
posi-
tioners are CSA certified and FM approved in the categories listed.
2. Hazardous location approvals for use in flammable atmospheres are for ambient conditions of
-25
°
to 40
°
C (-13
°
to 104
°
F), 86 to 106 kPa (12.5 to 15.7 psi) with a maximum oxygen concentra-
tion of 21%.