Electromagnetic Flowmeter
AquaMaster™ Explorer
3 Electrical Installation
IM/AM/E
Issue 10
23
3.4.6 Pressure Transducer (Optional)
Optional pressure transducer cables are available for a range of pressures and cable lengths.
Fig. 3.13 AquaMaster Explorer fitted with Optional Pressure Transducer Connector (Bulgin)
Fig. 3.14 AquaMaster Explorer fitted with Optional PressureTransducer Connector (MIL Style)
Caution.
Use only the pressure transducer supplied with the transmitter. Use of other pressure
transducers will require alteration of the pressure span and zero factors in the transmitter.