AquaMaster 3 
Electromagnetic flowmeter 

 

MODBUS Tables Supplement COI/FET2XX/MOD/TBL-EN 

 16 

 

MODBUS Protocol 

RO String 

Address

 

AquaMaster 3 

Parameter

 

Name

 

Size 
(bytes)

 

Note

 

2001

 

000

 

ApplicationVersionString

 

34

 

 

2019

 

017

 

FlowSensorContractNo

 

25

 

 

2037

 

018

 

FlowSensor1stFactory

 

11

 

Sensor first factory calibration date  
(see Table A.4 Date Format) 

2055

 

019

 

FlowSensorLastFactory

 

11

 

Sensor last factory calibration date  
(see Table A.4 Date Format)

 

2073

 

020

 

FlowSensorCertNo

 

25

 

 

2091

 

101

 

AmodemVersion

 

25

 

 

2109

 

199

 

TransmitterType

 

25

 

'AquaMaster 3'

 

2127

 

200

 

Manufacturer

 

25

 

 

2145

 

201

 

Address1

 

25

 

 

2163

 

202

 

Address2

 

25

 

 

2181

 

203

 

Address3

 

25

 

 

2199

 

204

 

Address4

 

25

 

 

2217

 

210

 

CalibrationDate

 

11

 

Transmitter calibration date  
(see Table A.4 Date Format)

 

2235

 

211

 

CertificateNo

 

25

 

 

2253

 

214

 

LoggerVersion

 

25

 

 

2271

 

215

 

OriginalManufacturer1

 

33

 

 

2289

 

216

 

OriginalManufacturer2

 

33

 

 

2307

 

280

 

FlowSensorString

 

25

 

 

2325

 

281

 

PressureSensorString

 

25

 

 

2343

 

282

 

TransmitterString

 

25

 

 

2361

 

285

 

ABB Reserved

 

25

 

 

2379

 

286

 

ABB Reserved

 

25

 

 

2397

 

365

 

ABB Reserved

 

25

 

 

2415

 

381

 

ABB Reserved

 

12

 

 

2433

 

383

 

ABB Reserved

 

12

 

 

2451

 

392

 

ABB Reserved

 

12

 

 

2469

 

408

 

BootProgId

 

34

 

 

2487

 

409

 

UamProgId

 

34

 

 

2505

 

415

 

ApplicationCRC

 

12

 

CRC value for MID

 

 

Summary of Contents for AquaMaster 3

Page 1: ...ation information and MODBUS Protocol Register Allocation tables and is applicable to product firmware version WAJC2103 v01 00 00 or above Detailed specifications and recommendations for using and imp...

Page 2: ...committed to teamwork high quality manufacturing advanced technology and unrivalled service and support The quality accuracy and performance of the Company s products result from over 100 years experi...

Page 3: ...ister Usage 4 1 3 1 RW 1 Bit 4 1 3 2 RO 1 Bit 5 1 3 3 RO 8 16 Bit 5 1 3 4 RO String 6 1 3 5 RO 32 Bit 7 1 3 6 RO Float 8 1 3 7 RW 8 16 Bit 8 1 3 8 RW String 10 1 3 9 RW 32 Bit 12 1 3 10 RW Float 13 Ap...

Page 4: ...dered when implementing a MODBUS master to communicate with AquaMaster 3 Transmission mode Remote Terminal Unit RTU slave Response time out 1s Turnaround delay 800ms Interface power up considerations...

Page 5: ...for each Register Address Range and Appendix A provides the Register allocation tables Register Address Range Applicable Function Codes AquaMaster Data Type Description 0001 0500 01 Read Coils 05 Writ...

Page 6: ...le bit types Read Coils 1 to 14 Read all RW 1 Bit items Request Message Hex Response Message Hex Error Message Hex Address 01 Address 01 Function 81 Function 01 Function 01 Exception Code Exception Co...

Page 7: ...Variable Meaning 0 501 0 007 Flow sensor is not reverse wired 1 502 1 008 Flow sensor is a Probe 2 503 1 416 MID read only switch is Active 3 to 7 Not Used 1 3 3 RO 8 16 Bit Read only access to 8 and...

Page 8: ...Each AquaMaster 3 String type requires 18 MODBUS Registers 36 bytes Consequently RO String Register addresses are assigned at intervals of 18 and the Request Message must provide Quantity of Inputs i...

Page 9: ...ters 4 bytes Consequently RO 32 Bit Register addresses are assigned at intervals of 2 and the Request Message must provide Quantity of Inputs in multiples of 2 32 Bit quantities are arranged in big en...

Page 10: ...g Address Hi 13 Byte count 04 Starting Address Lo a8 H Register Value Hi 41 Quantity of Inputs Hi 00 H Register Value Lo 20 Quantity of Inputs Lo 02 L Register Value Hi 00 CRC f4 L Register Value Lo 0...

Page 11: ...Value Hi 00 Register Value Hi 00 Register Value Lo 0e 14 Register Value Lo 0e 14 CRC ae CRC ae CRC 0b CRC 0b Write Multiple Registers 8019 var125 Trip point for high flow 15 Request Message Hex Respo...

Page 12: ...nse Message Hex char Error Message Hex Address 01 Address 01 Function 83 Function 03 Function 03 Exception Code Exception Code Starting Address Hi 27 Byte count 24 Starting Address Lo e8 byte0 41 A Qu...

Page 13: ...Address Lo d6 Starting Address Lo d6 Quantity of Registers Hi 00 Quantity of Registers Hi 00 Quantity of Registers Lo 12 Quantity of Registers Lo 12 Byte count 24 CRC ab byte0 46 F CRC 48 byte1 6c l...

Page 14: ...antity of Registers in multiples of 2 32 Bit quantities are arranged in big endian order Read Holding Registers 12003 to 12004 var224 Totaliser Forward Request Message Hex Response Message Hex Error M...

Page 15: ...s Hi 00 Register Value H Lo 66 Quantity of Inputs Lo 02 Register Value L Hi 66 CRC 0a Register Value L Lo 67 0 9000000 CRC 8c CRC 7c CRC 72 Each AquaMaster 3 RW Float type requires 2 MODBUS Registers...

Page 16: ...al on AquaMaster 3 LCD display 2 053 DisplayReverseTotal 0 Off 1 On 3 054 DisplayNetTotal 0 Off 1 On 4 055 DisplayTariffA 0 Off 1 On 5 056 DisplayTariffB 0 Off 1 On 6 059 DisplayFlowRate 0 Off 1 On 7...

Page 17: ...Sensor settling time mains power 1008 034 ABB Reserved 2 1009 035 ABB Reserved 2 1010 036 ABB Reserved 1 1011 106 CalibrationType 1 0 to 4 See Table A 2 Calibration Types 1012 148 ABB Reserved 1 1013...

Page 18: ...2109 199 TransmitterType 25 AquaMaster 3 2127 200 Manufacturer 25 2145 201 Address1 25 2163 202 Address2 25 2181 203 Address3 25 2199 204 Address4 25 2217 210 CalibrationDate 11 Transmitter calibrati...

Page 19: ...03 002 FlowSensorSoftwareRev 4 4005 003 FlowSensorHardwareRev 4 4007 172 PressureSoftwareRev 4 4009 173 PressureHardwareRev 4 4011 205 HardwareRevisionLevel 4 4013 207 TransmitterUniqueID 4 Transmitte...

Page 20: ...tedBore 4 mm 5035 238 ISpan 4 Factory coil current span factor 5037 239 IZero 4 Factory coil current zero correction 5039 243 CoilCurrent 4 A 5041 258 PulseOutput 4 Hz 5043 294 SignalCalHighCurrent 4...

Page 21: ...to 19 See Note 1 8016 117 FlowCutOffLow 1 0 to 255 mms 1 y Factory flow cut off battery mode 8017 118 UserFlowCutOff 1 0 to 255 y User s flow cut off setting 8018 119 PressureUnits 1 0 to 7 See Note...

Page 22: ...15 V 8054 326 ABB Reserved 1 8055 327 ABB Reserved 1 y 8056 334 SoftwareResetCount 2 8057 335 WatchdogResetCount 2 8058 336 SensorCalTxVersion 1 y 8059 339 ABB Reserved 2 8060 342 LastTxSoftwareVer 1...

Page 23: ...g 25 10271 186 Pressure1stFactory 11 See Table A 4 Date Format 10289 187 PressureLastFactory 11 See Table A 4 Date Format 10307 188 PressureCertNo 25 10325 189 PressureContractNo 25 10343 196 Pressure...

Page 24: ...itted unless MID active 12005 225 TotaliserRev 4 Totaliser units y Write value of 0 permitted unless MID active 12007 226 TotaliserNet 4 Totaliser units y Write value of 0 permitted unless MID active...

Page 25: ...15 110 CustomVelocityUnits 4 per ms 1 13017 113 CustomFlowUnits 4 per m3 s 1 13019 115 FlowUpperRange 4 Flow units y 13021 116 FlowLowerRange 4 Flow units y 13023 120 CustomPressureUnits 4 per mmHg 13...

Page 26: ...ttings 0 Special 1 N A 2 PFA 3 PTFE 4 FEP 5 ABR 6 Rubber WRC 7 Polyurethane 8 HT Rubber 9 Polypropylene 10 PES 11 P E 12 PPS 13 Viton 14 St St 15 C St 16 Brass 17 Gold 18 Titanium 19 Hast B 20 Hast C...

Page 27: ...OW 2 Battery Power FAIL 3 Mains Power ON 4 Mains Power OFF 5 Renewable Power OK 6 Renewable Power LOW 7 Renewable Power OFF Table A 3 External Power Status Modbus register 8030 AquaMaster 3 parameter...

Page 28: ...AquaMaster 3 Electromagnetic flowmeter MODBUS Tables Supplement COI FET2XX MOD TBL EN 26 MODBUS Protocol Notes...

Page 29: ...AquaMaster 3 Electromagnetic flowmeter MODBUS Tables Supplement COI FET2XX MOD TBL EN 27 MODBUS Protocol...

Page 30: ...AquaMaster 3 Electromagnetic flowmeter MODBUS Tables Supplement COI FET2XX MOD TBL EN 28 MODBUS Protocol...

Page 31: ...nalytics Instrumentation pH Conductivity and Dissolved Oxygen Transmitters and Sensors Ammonia Nitrate Phosphate Silica Sodium Chloride Fluoride Dissolved Oxygen and Hydrazine Analyzers Zirconia Oxyge...

Page 32: ...t without prior notice With regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in thi...

Reviews: