8
A DS 4 20
| D I SS O LV ED OX YG EN P R O B E | O I/A DS 4 2 0 - EN R E V. A
…3 Calibration
…1-Point calibration (water-saturated air)
9
If
1-Point Cal
is displayed, go to step 12.
If
1-Point Cal
is not displayed, press the
key
(below the
Edit
prompt).
The
Calibration Type
selection page is displayed:
Calibration Type
1-Point Cal
2-Point Cal
Cancel
OK
10
Use the
/
keys to select
1-Point Cal
and press the
key (below the
OK
prompt).
The
Calibration Type
page is displayed:
S1: DO
Next
Edit
Calibration Type
1-Point Cal
11
Press the
key below the
Next
prompt).
The
Calibrate/Start Span Cal?
page is displayed:
Calibrate
Start Span Cal?
Place in 100 % Sat
Abort
Continue
12
Press the
key (below the
Continue
prompt).
The
Calibrate
page is displayed:
Calibrate
PV 98.26 %
Tmp 25.22 °C
Setting-Please Wait
Abort
13
When span calibration is complete, a results page is
displayed:
Calibrate
PV 98.98 %
Tmp 24.90 °C
Offset 0.045
Slope 1.054
Exit
In the event of a calibration failure, the message
Calibration
Failed
is displayed with an indication of the reason for
failure, for example ‘
Result Out Of Bounds
’.
14
Press the
key (below the
Exit
prompt).
The
Calibrate page
is displayed:
Calibrate
S1: DO
Back
Select
15
Press the
key (below the
Back
prompt).
The
Calibrate page
is displayed:
Calibrate
Menu
Exit
Select
16
Press the
key (below the
Exit
prompt) to return to the
Operator page
.