ABB ADS420 Operating	 Instruction Download Page 8

8

A DS 4 20

  |  D I SS O LV ED OX YG EN P R O B E  |  O I/A DS 4 2 0 - EN R E V. A

…3  Calibration

…1-Point calibration (water-saturated air)

If 

1-Point Cal

 is displayed, go to step 12.

If 

1-Point Cal

 is not displayed, press the 

 key  

(below the 

Edit

 prompt).  

The 

Calibration Type

 selection page is displayed:

Calibration Type

1-Point Cal

2-Point Cal

Cancel

OK

10 

Use the 

/

 keys to select 

1-Point Cal

 and press the 

 

key (below the 

OK

 prompt). 

The 

Calibration Type

 page is displayed:

S1: DO

Next

Edit

Calibration Type

1-Point Cal

11 

Press the 

 key below the 

Next

 prompt). 

The 

Calibrate/Start Span Cal?

 page is displayed:

Calibrate

Start Span Cal?

Place in 100 % Sat

Abort

Continue

12 

Press the 

 key (below the 

Continue

 prompt). 

The 

Calibrate

 page is displayed:

Calibrate

PV      98.26 %
Tmp     25.22 °C

Setting-Please Wait

Abort

13 

When span calibration is complete, a results page is 
displayed:

Calibrate

PV      98.98 % 

Tmp     24.90 °C

Offset  0.045 

Slope   1.054

Exit

In the event of a calibration failure, the message 

Calibration 

Failed

 is displayed with an indication of the reason for 

failure, for example ‘

Result Out Of Bounds

’.

14 

Press the 

 key (below the 

Exit

 prompt). 

The 

Calibrate page

 is displayed:

Calibrate

S1: DO

Back

Select

15 

Press the 

 key (below the 

Back

 prompt). 

The 

Calibrate page

 is displayed:

Calibrate

Menu

Exit

Select

16 

Press the 

 key (below the 

Exit

 prompt) to return to the 

Operator page

.

Summary of Contents for ADS420

Page 1: ...ional EZClean compressed air supply unit connected to an ABB adaptor nozzle or T piece a schedule can be configured to provide automatic in situ cleaning of the probe For more information Further publ...

Page 2: ...cable attached 5 ADS420 cable detached 5 3 Calibration 6 Calibrate menu 6 1 Point calibration water saturated air 7 2 Point calibration 100 and 0 saturation 9 Abort calibration 11 Calibration timings...

Page 3: ...appear on this product are shown below Direct current supply only This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals shoul...

Page 4: ...ablish and maintain any appropriate measures such as but not limited to the installation of firewalls application of authentication measures encryption of data installation of anti virus programs etc...

Page 5: ...Note The sensor cap serial number is programmed on the memory chip inside the sensor cap D Probe body including temperature sensor E 10 m 39 3 ft fixed cable ADS420 cable detached ADS420 probe compone...

Page 6: ...After calibration and probe replacement reset the currents outputs and alarms see Calibrate menu below Calibrate menu Calibrate Menu Exit Select Used to calibrate the probe Access to the Calibrate me...

Page 7: ...roducing false low readings after calibration If condensation occurs remove the probe thoroughly dry the sensing element probe and thermistor before performing the calibration procedure Ensure the cal...

Page 8: ...displayed Calibrate Start Span Cal Place in 100 Sat Abort Continue 12 Press the key below the Continue prompt The Calibrate page is displayed Calibrate PV 98 26 Tmp 25 22 C Setting Please Wait Abort...

Page 9: ...he probe in the calibration chamber for more than 30 minutes this can allow condensation to form on the surface of the sensing element producing false low readings after calibration If condensation oc...

Page 10: ...Press the key below the Next prompt The Calibrate Start Span Cal page is displayed Calibrate Start Span Cal Place in 100 Sat Abort Continue 12 Press the key below the Continue prompt The Calibrate pag...

Page 11: ...nstructions OI AWT420 EN The Calibration Log displays the message Cal Aborted refer to OI AWT420 EN for Calibration Log entries Calibration timings Stability period The temperature and concentration r...

Page 12: ...re 3 Cleaning the sensor cap Cleaning the optical lens 1 Isolate the transmitter and remove the probe from the process Referring to Figure 4 2 Unscrew shroud A from probe body B 3 Grip sensor cap C be...

Page 13: ...low shows sensor specific icon types diagnostic messages and possible causes suggested remedial action Note The diagnostic icons in the following tables conform to NAMUR 107 For transmitter specific d...

Page 14: ...T4009010 AWT4009050 AWT4009100 AWT4009150 AWT4009250 AWT4009500 1 m 3 3 ft 5 m 16 4 ft 10 m 32 8 ft 15 m 49 2 ft 25 m 82 0 ft 50 m 164 0 ft Sensor cap spare parts Part number Description 3KXA494400L00...

Page 15: ...ADS420 DISSOLVED OX YGEN PROBE OI ADS420 EN REV A 15 Acknowledgements EZLink is a trademark of ABB Limited...

Page 16: ...h regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all...

Reviews: