ABB ADS420 Operating	 Instruction Download Page 3

A DS 4 20

  |  D I SS O LV ED OX YG EN P R O B E  |  O I/A DS 4 2 0 - EN R E V. A

3

1  Health & Safety

Document symbols

Symbols that appear in this document are explained below:

 WARNING

This symbol in conjunction with the signal word 'WARNING' 
indicates a potentially dangerous situation. Failure to 
observe this safety information may result in death or 
severe injury.

NOTICE  

The signal word ‘

NOTICE

’ indicates potential material 

damage.

‘Note’ 

indicates operator tips, particularly useful information or 

important information about the product or its further uses. 

‘Note’

 does not indicate a dangerous or harmful situation.

Safety precautions

Be sure to read, understand and follow the instructions 
contained within this manual before and during use of the 
equipment. Failure to do so could result in bodily harm or 
damage to the equipment.

 WARNING

Installation, operation, maintenance and servicing must be 
performed:

•  by suitably trained personnel only

•  in accordance with the information provided in this 

manual

•  in accordance with relevant local regulations

Potential safety hazards

ADS420 probe – electrical

The probe operates on 24 V DC supplied from the transmitter.
There are no hazardous voltages present.

Safety standards

This product has been designed to satisfy the requirements of
IEC61010-1:2010 3rd edition 'Safety Requirements for
Electrical Equipment for Measurement, Control and Laboratory
Use' and complies with US NEC 500, NIST and OSHA.

Product symbols

Symbols that may appear on this product are shown below:

  Direct current supply only.

  This symbol identifies a risk of chemical harm and 

indicates that only individuals qualified and trained to 
work with chemicals should handle chemicals or 
perform maintenance on chemical delivery systems 
associated with the equipment.

  This symbol indicates the need for protective eye 

wear.

  This symbol indicates the need for protective hand 

wear.

  Recycle separately from general waste under the 

WEEE directive.

Product recycling and disposal (Europe only)

  ABB is committed to ensuring that the risk of any 

environmental damage or pollution caused by any of 
its products is minimized as far as possible. The 
European Waste Electrical and Electronic Equipment 
(WEEE) Directive that initially came into force on 
August 13 2005 aims to reduce the waste arising from 
electrical and electronic equipment; and improve the 
environmental performance of all those involved in 
the life cycle of electrical and electronic equipment. In 
conformity with European local and national 
regulations, electrical equipment marked with the 
above symbol may not be disposed of in European 
public disposal systems after 12 August 2005.

NOTICE 

For return for recycling, please contact the equipment 
manufacturer or supplier for instructions on how to return 
end-of-life equipment for proper disposal.

Information on RoHS Directive 2011/65/EU 

(RoHS II)

  ABB, Process Automation, Measurement & Analytics, 

UK, fully supports the objectives of the RoHS II 
directive.  All in-scope products placed on the market 
by PAMA UK on and following the 22

nd

 of July 2017, will 

be compliant to the RoHS directive, 2011/65/EU.

Summary of Contents for ADS420

Page 1: ...ional EZClean compressed air supply unit connected to an ABB adaptor nozzle or T piece a schedule can be configured to provide automatic in situ cleaning of the probe For more information Further publ...

Page 2: ...cable attached 5 ADS420 cable detached 5 3 Calibration 6 Calibrate menu 6 1 Point calibration water saturated air 7 2 Point calibration 100 and 0 saturation 9 Abort calibration 11 Calibration timings...

Page 3: ...appear on this product are shown below Direct current supply only This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals shoul...

Page 4: ...ablish and maintain any appropriate measures such as but not limited to the installation of firewalls application of authentication measures encryption of data installation of anti virus programs etc...

Page 5: ...Note The sensor cap serial number is programmed on the memory chip inside the sensor cap D Probe body including temperature sensor E 10 m 39 3 ft fixed cable ADS420 cable detached ADS420 probe compone...

Page 6: ...After calibration and probe replacement reset the currents outputs and alarms see Calibrate menu below Calibrate menu Calibrate Menu Exit Select Used to calibrate the probe Access to the Calibrate me...

Page 7: ...roducing false low readings after calibration If condensation occurs remove the probe thoroughly dry the sensing element probe and thermistor before performing the calibration procedure Ensure the cal...

Page 8: ...displayed Calibrate Start Span Cal Place in 100 Sat Abort Continue 12 Press the key below the Continue prompt The Calibrate page is displayed Calibrate PV 98 26 Tmp 25 22 C Setting Please Wait Abort...

Page 9: ...he probe in the calibration chamber for more than 30 minutes this can allow condensation to form on the surface of the sensing element producing false low readings after calibration If condensation oc...

Page 10: ...Press the key below the Next prompt The Calibrate Start Span Cal page is displayed Calibrate Start Span Cal Place in 100 Sat Abort Continue 12 Press the key below the Continue prompt The Calibrate pag...

Page 11: ...nstructions OI AWT420 EN The Calibration Log displays the message Cal Aborted refer to OI AWT420 EN for Calibration Log entries Calibration timings Stability period The temperature and concentration r...

Page 12: ...re 3 Cleaning the sensor cap Cleaning the optical lens 1 Isolate the transmitter and remove the probe from the process Referring to Figure 4 2 Unscrew shroud A from probe body B 3 Grip sensor cap C be...

Page 13: ...low shows sensor specific icon types diagnostic messages and possible causes suggested remedial action Note The diagnostic icons in the following tables conform to NAMUR 107 For transmitter specific d...

Page 14: ...T4009010 AWT4009050 AWT4009100 AWT4009150 AWT4009250 AWT4009500 1 m 3 3 ft 5 m 16 4 ft 10 m 32 8 ft 15 m 49 2 ft 25 m 82 0 ft 50 m 164 0 ft Sensor cap spare parts Part number Description 3KXA494400L00...

Page 15: ...ADS420 DISSOLVED OX YGEN PROBE OI ADS420 EN REV A 15 Acknowledgements EZLink is a trademark of ABB Limited...

Page 16: ...h regard to purchase orders the agreed particulars shall prevail ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document We reserve all...

Reviews: