A DS 4 20
| D I SS O LV ED OX YG EN SEN S O R | O I/A DS 4 2 0 - EN R E V. B
7
4 Calibration
This section describes how to calibrate the sensor and involves
measuring the sensor's sensitivity to oxygen by exposing the
sensor to water-saturated air.
Calibrations are initiated via the
Cal
prompt displayed on
Operator pages or via the
Calibrate
and
Advanced
menu
items on the
Access Level
page – refer to transmitter Operating
instructions
for all transmitter menu options.
Note .
• Do not perform a calibration until the sensor and transmitter
are installed and ready for operation.
• Before removing the sensor for calibration purposes, set the
currents outputs and alarms to Hold (enabled via the
Operator Menu/Manual Hold function).
• After calibration and sensor replacement, reset the currents
outputs and alarms – see
below.
Calibrate menu
Calibrate
Menu
Exit
Select
Used to calibrate the sensor.
Access to the Calibrate menu is permitted via the
Calibrate
and
Advanced
levels only.
Note .
During calibration, current outputs and alarms are set to
Hold
automatically if
Hold Outputs
is enabled (see below).
Menu
Comment
Default
S1 (to 4) :DO
Select the optical dissolved oxygen sensor to calibrate.
Probe Calibration
Calibration Type
1-Point Cal
Select to perform a 1-point calibration in water-saturated air (100% saturation):
2-Point Cal
Select to perform a 100 % saturation calibration in water-saturated air and a 0 % saturation
calibration in an oxygen-depleted solution, e.g. sodium sulfite (Na
2
SO
3
).
Hold Outputs
Enable/disable the
Hold Outputs
function. If enabled, the current outputs and alarm functions are
held during calibrations.
Enabled