8

Install a suitable solenoid valve to

an appropriate water supply.

1

Connect a suitable length of

1

/

4

" i.d. reinforced p.v.c. tube

to the outlet of the solenoid

valve; secure with a clamp or

clip.

Push the quick-fit

connector supplied onto

the free end of the tube;

secure with a clamp or clip.

Note that the quick-fit

connector incorporates a

cut-off valve.

To water supply – see

Important Note and

Caution above.

Suitable Solenoid Valve (user supplied)

Plug the quick-fit connector

onto the mating coupling

fitted to the sensing system

incorporating the water wash

option.

Note. Allow enough tubing to permit handling of the

system in and out of the sample. Alternatively,

disconnect tube at the quick-fit connector.

2

3

4

3.3  Water Supply for Water Wash Systems – Fig. 3.6

Important Note. Installation must only be carried out in 

accordance with the local water authority and council 

bylaws.

Caution. The maximum water pressure at the electrode 

should not exceed  50 psi (4 bar). At NO time should the 

sample pressure be allowed to exceed that of the water 

wash water supply. Fit a non return valve if this possibility 

exists.

Information.  As  all  quick-fit  connectors  used  in  Water 

Wash  systems  incorporate  cutoff  valves,  tubes  may 

subsequently be disconnected on operational equipment 

without sample loss or water wash discharge.

Note. For optimum performance of the Water Wash 

function in flow systems, the sample pressure should be 

at least 2 bars lower than that of the water wash water 

supply.

The water supply for the water wash system must come from 

a suitable supply via a solenoid valve. A general specification 

for the solenoid valve is:

Brass body: .........2/2, NC

Orifice size: .........3.0 / 4.0 mm

Port size: .............

1

/

4

 in – 

1

/

2

 in BSP or NPT

Pressure: .............0 to 6 bar

Coil: .....................24 V a.c.,  110/115 V a.c.  or

 

230/240 V a.c. 50/60 Hz

Note. See Fig. 3.6 for recommended tubing.

Fig. 3.6 Installing the Water Wash Water Supply

…3  MECHANICAL INSTALLATION

Summary of Contents for 9408 Series

Page 1: ...xygen system comprises a flat ended sensor mounted into a relatively large diameter spherical flotation collar This form of construction presents to the sample a surface on which it is difficult for rags and other large solids to lodge For more demanding applications on line cleaning is available using the water wash option 9408 series Dissolved oxygen systems ...

Page 2: ... more information Further publications are available for free download from www abb com analytical or by scanning this code Search for or click on 9408 series Dissolved oxygen systems Data Sheet DS 9408 EN ...

Page 3: ...npacking 2 2 2 Floating Ball System 2 2 2 1 Assembled Version 2 2 2 2 Kit Version 2 2 3 Dip Systems 3 2 3 1 Assembled Version 3 2 3 2 Kit Version 3 2 4 Submersible System 3 3 MECHANICAL INSTALLATION 4 3 1 Assembling Kit Systems 4 3 1 1 Floating Ball 4 3 1 2 Dip 5 3 2 Installation 6 3 2 1 Floating Ball System 6 3 2 2 Dip Systems 7 3 2 3 Submersible Systems 7 3 3 Water Supply for Water Wash Systems ...

Page 4: ...I T P O M E T S Y S d r a d n a t S r e t a w h t i W h s a W h s a W h s a W h s a W h s a W c i r t e M 0 5 7 8 0 4 9 2 5 7 8 0 4 9 l a i r e p m I 0 6 7 8 0 4 9 2 6 7 8 0 4 9 s n o i s n e m i D e p i P c i r t e M 0 1 N P t w 3 3 x m m 0 5 d o e p i p S B A l a i r e p m I E s s a l C t w 5 4 x B N 2 1 1 e p i p S B A 2 1 Unpacking After carefully removing the 9408 system from its packing chec...

Page 5: ...o scale Ø40 mm User supplied tube ABS or PVC 1 9 inches O D Contents 1 Top cap with fixing screw 2 Adaptor 3 Shroud 4 Sensor block with cable attached and dummy sensor fitted 5 Dissovled oxygen sensor capsule 6 Tangit Solvent Cement not provided Contents 1 Top cap with fixing screw 2 Adaptor 3 Shroud 4 Sensor block with cable and water wash tube attached and dummy sensor fitted 5 Dissovled oxygen ...

Page 6: ...4 4 5 4 3 1 Assembling Kit Systems 3 1 1 Floating Ball Fig 3 1 For installation of the assembled unit see Section 3 2 1 Floating Ball System Fig 3 1 Floating Ball Kit Assembly 3 MECHANICAL INSTALLATION ...

Page 7: ...5 3 MECHANICAL INSTALLATION 3 1 2 Dip Fig 3 2 For installation of the assembled unit see Section 3 2 2 Dip Systems Fig 3 2 Dip Kit Assembly ...

Page 8: ...nt cable slack for positioning servicing procedure Make the necessary electrical connections see Section 4 ELECTRICAL INSTALLATION If applicable arrange to have a water wash mains supply available see Section 3 3 and connect to the quick fit connector on the end cap 1 3 2 5 6 4 Standard Swivel Standard Swivel C Standard Swivel C 30 Viewed from above 3 2 Installation Caution The system is supplied ...

Page 9: ...rovided This is to prevent cable strain and allows the system to be lowered to the required depth 3 2 2 Dip Systems Fig 3 4 Note The standard dip system is supplied with the temperature probe and oxygen sensor body fitted Before fitting the dip tube to any bracket or fixing remove the protective shroud unscrew the dummy capsule from the end of the oxygen sensor body and fit the oxygen sensor capsu...

Page 10: ...cal water authority and council bylaws Caution The maximum water pressure at the electrode should not exceed 50psi 4 bar At NO time should the sample pressure be allowed to exceed that of the water wash water supply Fit a non return valve if this possibility exists Information As all quick fit connectors used in Water Wash systems incorporate cutoff valves tubes may subsequently be disconnected on...

Page 11: ...ively 4 2 Calibrating the Sensor Fig 4 1 The procedures for zero and span calibration are fully described in the 4600 Series Transmitter Unit Operating Instructions to which reference should be made Note It is advisable to switch off the water wash before starting the calibration procedure This can be done either by turning off at the 4600 transmitter unit or by disconnecting the quick fit connect...

Page 12: ... Section 4 2 Calibrating the Sensor Caution Do not leave the new capsule exposed to air for more than 30 minutes as the membrane will dry out Ensure the following steps are done carefully to avoid damaging the membrane covering the silver cathode Align the holes in the end of the ball with the protruding sensor capsule and temperature compensator and replace the ball assembly over the sensor pipe ...

Page 13: ...n box if applicable and attach a draw wire of suitable length 3 Slacken the cable gland on the end of the boom 4 Remove the self tap screw and pull off the end cap 5 Draw the cable through the gland 6 Unscrew the securing ring of the ball float whilst restraining the ball from turning and withdraw the ball from the elbow union 7 Carefully unscrew the sensor block from the end of the sensor pipe 8 ...

Page 14: ...ure Compensation Automatic correction by means of integral Pt100 resistance thermometer Accuracy 0 2mgl 1 or 2 saturation within 10 C of the calibration temperature in the range 0 to 35 C Operating Temperature Range 0 to 40 C Operating Pressure Atmospheric Dimensions Floating Ball Ball diameter 203 mm Boom outside diameter Metric 50 mm Imperial 17 8 in 48mm approximately Boom length Assembled 3 me...

Page 15: ...l System Table 7 1Floating Ball Spares List Fig 7 2Dip System Table 7 2Dip Spares List 7 SPARE PARTS LIST m e t I o N n o i t p i r c s e D o N t r a P 1 y l b m e s s A t a o l F l l a B 5 6 0 8 0 4 9 2 t e k c a r B d r a d n a t S y l b m e s s A c i r t e M 5 1 1 8 0 4 9 l a i r e p m I 6 1 1 8 0 4 9 A 2 t i K y l b m e s s A l e v i w S c i r t e M 5 3 1 8 0 4 9 l a i r e p m I 6 3 1 8 0 4 9 ...

Page 16: ...rsion Kit Fig A 1 Items in Conversion Kit Fig A 2 Assembled Kit APPENDIX A SUPPORT BRACKET CONVERSION m e t I o N o N o N o N o N n o i t p i r c s e D o N f f O f f O f f O f f O f f O s o N t P 1 s p m a l c m o o B 4 t i K c i r t e M 5 3 1 8 0 4 9 t i K l a i r e p m I 6 3 1 8 0 4 9 2 s d n e s p m a l c m o o B 2 3 s t l o b s p m a l c m o o B 4 4 w e r c s r e d l u o h S 1 5 e t a l p l e ...

Page 17: ...15 NOTES ...

Page 18: ...16 NOTES ...

Page 19: ...Sales Service ...

Page 20: ...ights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB ABB 2018 ABB Limited Measurement Analytics Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel 44 0 1453 826661 Fax 44 0 1453 829671 Email instrumentation gb abb co...

Reviews: