32
Return to
top of page
.
Monitoring pH
7 0 0
Redox (ORP)
2 5 0
or
Alter Fact. Code
0 0 0 0 0
Adjust RTX Zero2
– – – – –
Adjust RTX Span2
– – – – –
Continued from previous page.
Adjust RTX Zero
– – – – –
Adjust RTX Span
– – – – –
…8
CALIBRATION
…8.3
Factory Settings Page
Adjust Retransmission Zero
Set the milliammeter reading to 4mA.
✶
Note. The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Advance to next parameter.
Adjust Retransmission Span
Set the milliammeter reading to 20mA.
✶
Note. The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Zero (available only on 4630/800, 4631/800, 4635/800 and
4636/800 instruments)
Set the milliammeter reading to 4mA.
✶
Note. The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Span (available only on 4630/800, and 4631/800, 4635/800
and 4636/800 instruments)
Set the milliammeter reading to 20mA.
✶
Note. The retransmission range selected in the
Set Up Retransmission Page
does
not affect the reading.
Alter Factory Settings Access Code
Set the factory settings access code to a value between 00000 and 19999.
Return to
Operating Page
.