Commissioning
40 265Dx,
265VS
IM/265D/V-EN-02
Pos: 15.12 /Überschriften/1.1/1-spaltig/D - F/Drehung des Gehäuses gegenüber dem Messwerk @ 5\mod_1165484889437_3101.doc @ 50488
6.4
Turning the housing in relation to the measuring equipment
Pos: 15.13 /Inbetriebnahme/Druck/Druckmessumformer/Allgemein/Drehung des Gehäuses gegenüber dem Messwerk @ 2\mod_1149843345046_3101.doc @ 29138
The electronics housing can be rotated 360 ° and fixed in any position; a stop is provided to
prevent it from rotating past this point.
• To activate this, slacken the housing stop screw on the neck of the housing (hex-head
socket screw SW 2.5) by approx. 1 rotation (do not pull it out) and, once the desired position
has been reached, retighten it until hand-tight.
Pos: 15.14 /Überschriften/1.1/1-spaltig/M - O/Montage / Demontage der Tasteinheit @ 2\mod_1149843645359_3101.doc @ 29159
6.5
Installing/Removing the button unit
Pos: 15.15 /Inbetriebnahme/Druck/Druckmessumformer/Allgemein/Montage / Demontage der Tasteinheit @ 2\mod_1149843760359_3101.doc @ 29180
1.
Slacken the screw on the protective cap and move the cap to one side.
2.
Using a suitable screwdriver, for example, push the lock bar all the way out of the button
unit.
3.
This releases the square nut; remove this from the button unit.
4.
Use a Torx screwdriver (size T10) to slacken the fixing screw for the button unit, and then
remove the unit from the electronics housing.
5.
If necessary, insert a filler piece and secure it using the screw supplied.
M00042
1
2
Fig. 14: Installing/removing the button unit
1
Fixing screw
2
Lock bar
Important
The fixing screw is located underneath the button unit.
Pos: 15.16 /======= Seitenumbruch ======== @ 0\mod_1126532365768_3101.doc @ 3830