Cation Exchange Column
2155
5 SPECIFICATION
IM/2155 Rev. G
5
5 SPECIFICATION
5.1 General
5.2 Handling and Storage Conditions for Columns
and Resin
Store between 5 ° and 50 °C (41 ° and 122 °F).
Avoid freezing.
Avoid de-hydration.
When handling the resin use of gloves and safety glasses is
recommended.
Fig. 4.1 Partially Depleted Column
Depleted
Resin
Maximum Flow rate:
400 ml min
-1
Minimum Flow Rate:
30 ml min
-1
Recommended Flow Rate:
100 to 300 ml min
-1
Max Temperature:
50 °C
Max Pressure:
3 kgcm
-1
/ 3 bar / 45 psi
Ion Exchange ADH Resin:
Colour changes when
exhausted from orange to
dark red
Connection to Inlet:
6 mm (
1
/
4
in.) ID
Connection to Outlet:
6 mm (
1
/
4
in.) ID
Caution.
Spilled resin beads can cause a slip hazard.