Vetscan HM5 Reagent Pack

 

Installation Guide & Package Insert

Technical Support Phone Number 1-800-822-2947
©January 2010, Abaxis Inc., Union City, CA 94587 

 

 

 

 

 

 

Abaxis Part No. 770-9000

INTENDED USE:  the VetScan HM5 reagent pack is intended for 
veterinary hematology use with the VetScan HM5 analyzer.

COMPONENTS INCLUDED:  

Diluent (green cap)

 

†

Lyse (yellow cap)

 

†

Rinse (white cap)

 

†

Lyse2 (white/orange cap)

 

†

Cleaner (blue cap)

 

†

Thermal paper roll (one)

 

†

STORAGE:  

Store reagent pack at room temperature (68-86° F).  Do not use after  
the expiration date listed on top of carton. Do not freeze reagents.   
Do not use if reagents have been previously frozen. Reagents are  
stable for 26 weeks after opening, if stored at optimum conditions. 

Allow your reagent pack to equilibrate to room temperature for 24 hours 
prior to use to prevent the formation of micro-bubbles that may affect 
your results.

CHANGING THE REAGENT PACK

Before changing the reagent pack, be sure to read the instructions 
carefully. You will need the following:

Distilled water (approximately 100-200 mls, or  

 

†

about 1 cup) – do not use tap water

5 ml of household bleach  

 

†

(approximately 5-6% hypochlorite)

HM5 cleaning tube

 

†

HM5 reagent pack

 

†

Two clean containers/cups to hold the water  

 

†

and bleach

HM5 Normal control (Abaxis part no. 770-9029)

 

†

Prepare a 1:20 dilution of bleach. 

1. 

 Mix 5 ml household bleach with 95 ml of 

a. 

distilled water.  

 Perform a bleach cleaning (

2. 

Utilities

 menu 

 , 

Maintenance

 

 Bleaching

), see HM5 User’s 

Manual, section 7.3 for details.

Reconnect the reagent tubing to the analyzer.  

3. 

 Unscrew the reagent caps from the old reagent 

4. 

pack. Connect them to the new reagent bottles.  

NOTE: 

Take care not to touch the drop down 

tubing (tubing that is inside the bottles) or the 
reagents may become contaminated. Be sure 
to match the correct caps to the color-matched 
bottles.

Press 

5. 

New

 for new reagent pack.  

 The software will ask if you are changing the 

a. 

reagent pack, or using an old one.  

Prime the reagent lines:  follow the menus below:

6. 

 Utilities 

a. 

key – 

Maintenance 

 Priming 

 Prime 

All

 Reset the Reagent Status:  follow the menus 

7. 

below:

 Utilities 

a. 

key – 

Maintenance 

 Reagent Status

 

 Press 

b. 

Change

 and the settings will be reset to 

100%. Press 

Accept

.

 If you have moved the reagent pack’s location 

8. 

from the previous pack’s location, you will need to 
reset the reagent sensors.  Follow the menu below:

 Utilities

a. 

 

 Settings 

 Fluid Sensors 

 Calibrate 

Sensors

Run a control sample.  

9. 

 Be sure to follow the procedure in the HM5 

a. 

User’s manual when running a control. Re-
calibrate the instrument if the control results 
do not match the expected values.

Your instrument is ready to go!

10. 

Doc. Number & Rev 770-7017 Rev B