July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
Page 63
: AADI Real-Time
Collector uses a default setting that fits
for most Smart Sensor. However the
DCPS sensor outputs a large amount of
data and might have much longer
response time (depending on the
configuration) than other smart sensors.
Some of the connection settings might
need to be changed. We recommend
using the settings as shown in
After updating the
Advanced
Connection Settings
, click on
Apply
and
OK
and then
OK
to go back to the
start screen.
Figure 3-5: Advanced connection setting
Press
Open Port
and the connection to
the SeaGuardII/SmartGuard should be
established within a few seconds and
the status turn to green. After power up
it will take approximately 30 seconds
before the logger is ready.
Open
Control Panel
in the lower right
).
Figure 3-6: AADI Real-Time Collector main menu
3.3.4
Control Panel
In the
Control Panel
window you will find four tabs,
Recorder Panel, Device Configuration Device Layout
and
System Status
. Under
Recorder Panel
you can start and stop recordings. If the recorder is running first click on the
Stop Recorder
as you are not allowed to configure the sensor when recording. For information about
Device Layout
and
System Status
see Manual for SeaGuardII Platform TD303 or SmartGuard TD293.