background image

 

English 5 

 

Features 

 

The UT-601 Ear Thermometer is a multi-use thermometer. It detects body 
temperature from the ear canal. It can also detect the temperature of solids, 
liquids or air in a fast and non-contact manner. 
The thermometer measures the infrared heat generated by the eardrum or 
other objects. Ear temperature accurately reflects core body temperature 
since the eardrum shares blood supply with the temperature control center in 
the brain (the hypothalamus).   
To ensure accuracy, the thermometer scans continuously while the 

 On/Scan 

   

button is being pressed and displays the highest temperature. 

 
 

  0 °C to 100 °C (32 °F to 212 °F) detection 

 

The display temperature range is from 0 °C to 100 °C. 

  So, besides measuring body temperature from the ear canal, the 

thermometer can be used in various ways. Examples of use are: 

 

To measure the temperature of milk or baby bath water. 

 

To measure the temperature of food or soup. 

 

To measure the temperature of solids, liquids or air. 

 

  °C or °F can be selected. 

 

  Easy one button operation 

 

  No probe cover is required during measurement. Wipe the probe clean 

with a soft cloth moistened with alcohol after use. 

 

  The last reading is displayed after power on. 

 

  Continuous scanning to ensure accuracy 

  The thermometer scans continuously while the 

 On/Scan 

  button is being 

pressed and displays the highest temperature. 

 

  Automatic power-shut off function 

The thermometer will turn off automatically after 30 seconds of no 
operation or after continuously pressing the 

 On/Scan 

 button for 30 

seconds. 

 

Summary of Contents for UT-601

Page 1: ...Ear Thermometer Model UT 601 Instruction Manual Original Manuel d instructions Manual de Instrucciones Manuale di Istruzioni 使用手冊 1WMPD4002420 ...

Page 2: ... Symbols 4 Features 5 Using the Thermometer 6 Installing Changing the Batteries 6 Measuring Ear Temperature 7 Measuring the Temperature of Liquid or Solid Objects 8 Measuring Air Temperature 8 Troubleshooting 9 Errors 9 Maintenance 10 Technical Data 10 Calibration Checks Testing Instructions 11 ...

Page 3: ... resistant Prevent rain sweat and water from soiling the thermometer Measurements may be distorted if the thermometer is used close to televisions microwave ovens cellular telephones X ray or other devices with strong electrical fields Keep the thermometer out of reach of children To ensure accuracy keep the lens of the probe clean No probe cover is required when using the thermometer For sanitary...

Page 4: ...tion Detection probe LCD On Scan button with mark Probe cover Battery cover Battery cover latch Part of Display Low battery mark MEMORY mark Temperature reading Unit of temperature Select before measurement Scanning in progress ...

Page 5: ...sure Replace both batteries with new ones when possible When flashes The battery is low Unable to measure Replace both batteries with new ones The ambient temperature is over 40 C The ambient temperature is below 5 C Use the thermometer in a 5 to 40 C environment Decimal point flashes The reading is beyond a body temperature range of 34 C to 42 2 C Try the measurement again The reading is over 100...

Page 6: ...perature range is from 0 C to 100 C So besides measuring body temperature from the ear canal the thermometer can be used in various ways Examples of use are To measure the temperature of milk or baby bath water To measure the temperature of food or soup To measure the temperature of solids liquids or air C or F can be selected Easy one button operation No probe cover is required during measurement...

Page 7: ... batteries and the poles CAUTION Insert the batteries as indicated into the battery compartment If not the device will not work When Low battery mark appears on the display it indicates that the battery is low At this stage measurement is still possible When Low battery mark flashes on the display replace both batteries with new ones Do not mix old and new batteries It may shorten the battery life...

Page 8: ...fter 30 seconds of no operation 3 The last reading stored in MEMORY appears and a short beep sounds to indicate that the thermometer is ready to scan 4 Insert the probe into the ear canal For babies gently pull the ear back to straighten the ear canal for easier insertion 5 Press and hold the On Scan button until a beep sounds 6 Release the On Scan button A long beep sounds to indicate completion ...

Page 9: ...etion of the temperature measurement Read the value Measuring Air Temperature Note The operating ambient temperature is 5 C to 40 C 1 Press the On Scan button to turn the thermometer on All the display segments appear 2 The last reading stored in MEMORY appears and a short beep sounds to indicate that the thermometer is ready to scan 3 Aim the probe toward any surrounding space with no objects nea...

Page 10: ...is over 40 C The ambient temperature is below 5 C Use the thermometer in a 5 C to 40 C environment Decimal point flashes The reading is beyond a body temperature range of 34 C to 42 2 C Try the measurement again The reading is over 100 C The reading is below 0 C Do not use the thermometer to measure the temperature of the object When appears The battery is low Still able to measure Replace both ba...

Page 11: ...emperature range 0 C to 100 C 32 F to 212 F Display resolution 0 1 C or 0 1 F Measurement accuracy In an ambient temperature of 16 C to 35 C 34 0 C to 35 4 C 0 3 C 35 5 C to 42 0 C 0 2 C 42 1 C to 42 2 C 0 3 C Others 1 0 C Power supply 3VDC Two alkaline button batteries LR44 AG13 Classification Type B EMC EN60601 1 2 2007 Memory Last reading Operating ambient temperature 5 C to 40 C 41 0 F to 104 ...

Page 12: ...y and deviatioin of pressure display Note Perform measurements in a 5 C to 40 C environment If the ambient temperature is not in the above temperature range an error or appears The display temperature range is 0 C to 100 C If the temperature of the object is out of the range an error or appears Testing Procedure 1 Press the On Scan button to turn the thermometer on 2 Place the thermometer on the s...

Page 13: ...A Telephone 1 408 263 5333 Fax 1 408 263 0119 A D Australasia Pty Ltd 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 A D KOREA Limited 한국에이 엔 디 주 대한민국 서울시 영등포구 여의도동 36 2 맨하탄 B D 8층 8th Floor Manhattan Bldg 36 2 Yoido dong Youngdeungpo ku Seoul KOREA 전화 82 2 780 4101 팩스 82 2 782 4280 A D RUS CO LTD Компания ЭЙ энд ДИ РУС 121357 Российская Федераци...