Calibration Checks-Testing Instructions
For trained specialist staff only
Calibration checks must be performed by the following institutioins or
persons:
Manufacturer
Trade measurements authorities
Persons who fulfil the requirements of the legislation governing the
operation of medical devices.
(A) Function testing
Function testing of the thermometer can only be carried out on a person
or with a suitable simulator.
(B) Testing of temperature circuit integrity and deviatioin of
pressure display
Note
Perform measurements in a 5 °C to 40 °C environment.
If the ambient temperature is not in the above temperature range, an error,
or
, appears.
The display temperature range is 0 °C to 100 °C.
If the temperature of the object is out of the range, an error,
or
,
appears.
Testing Procedure
1. Press the
On/Scan
button to turn the thermometer on.
2. Place the thermometer on the standard temperature generator such as a
blackbody radiation source.
3. Press and hold the
On/Scan
button until a beep sounds.
4. Release the
On/Scan
button.
The measured temperature is displayed.
5. The thermometer will turn off automatically after 30 seconds of no
operation.