English 12
Maintenance
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem using
the troubleshooting instructions, request service from your dealer or from the
A&D service group. The A&D service group will provide technical information,
spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life. However
it is generally recommended to have the monitor inspected every 2 years, to
ensure proper functioning and accuracy. Please contact either your
authorized dealer or A&D for maintenance.
Technical Data
Type UA-705
Measurement method
Oscillometric measurement
Measurement range
Pressure: 20 - 280 mmHg
Pulse:
40 - 200 beats / minute
Measurement accuracy
Pressure: ±3 mmHg or 2%, whichever is greater
Pulse:
±5%
Power supply
1 x 1.5V batteries (R6P, LR6 or AA)
Classification Type
BF
Clinical test
According to ANSI / AAMI SP-10 1987
EMC
IEC 60601-1-2: 2001
Data memory
Last 30 measurements
Operating condition
+10°C to +40°C / 30%RH to 85 %RH
Storage condition
-10°C to +60°C / 30%RH to 85 %RH
Dimensions
Approx. 81 [W] x 54 [H] x 105 [D] mm
Weight
Approx. 120 g, excluding battery
Note: Specifications are subject to change for improvement without prior
notice.