ENGLISH 3
Symbols
Function / Meaning
Treatment
Turns the device on and off.
___________
Guide to install battery
___________
Direct current
___________
SN
Serial number
___________
Date of manufacture
___________
Type BF: Device, cuff and tubing
are designed to provide special
protection against electrical
shocks.
___________
Appears when measurement is in
progress and flashes when the
pulse is detected.
Measurement is in progress -
- remain still.
Flashes when there is air
remaining in the cuff at power up.
Flashes at the completion of the
measurement until exhaust is
complete.
Press the exhaust valve
button to exhaust the air.
The pulse is detected immediately
after the measurement has begun.
Increase pressurization by 30
to 40 mmHg and measure
again.
Appears when the battery voltage
is too low for the device to work
properly.
Replace all batteries with
new ones.
Appears when the blood pressure
can not be measured accurately.
First exhaust the air from the
cuff, then re-apply the cuff
properly and measure again.
Appears when the pulse is not
detected correctly.
First exhaust the air from the
cuff, then re-apply the cuff
properly and measure again.
Flashing
exhaust
velosity
The exhaust rate is greater than 8
mmHg/s.
Adjust the exhaust rate to 2
to 5 mmHg/s.
The medical device label by the
EC directive
___________
Registered with the Australian
Communication Authority by the
Trademarks office.
___________