6.5. Pre-inspection
■
Perform the pre-inspection every day to ensure safe and correct usage.
6.5.1.
Introduction
Before using the monitor for the first time each day, perform the following pre-inspection.
6.5.2.
Before switching the power on
Is there any external deformation or damage to the monitor?
Is the monitor wet?
Is the monitor in a stable location free of tilting, vibrations and impacts?
Blood pressure measurement section
Is there damage or abnormalities around the arm insertion section (cuff area)?
Is the arm cuff cover attached?
Is the arm cuff cover overstretched?
Connection cable
Are the optional cables inserted firmly into the connectors of the monitor?
Power cable
Make sure that the electrical outlet is properly grounded and supplies the specified voltage
and frequency (100 V-240 V~ 50 Hz-60 Hz).
6.5.3.
After switching the power
on
Is there any smoke or strange smell?
Can you hear any strange noises?
Checking the time
Is the time set correctly?
If the time is incorrect when recording data, the data will be incorrect.
Checking the display
After switching the power on, all LEDs switch on for
several seconds and then blood pressure
measurement is possible. At this time, the diastolic
blood pressure display displays “0”.
Warning
DIA mmHg