SPECIFICATIONS
4.1. Model configuration
Model
Included functions
TM-2657P-EX
TM-2657P-EG
Printer
○
○
Measurement status LED
○
○
Time,Date format
24hour,DD/month/YYYY
12hour,month/DD/YYYY
4.2. Performance specifications
General
AC Power supply
100 V-240 V
~
50 Hz-60 Hz
Power consumption
50-80 VA
Safety standard
IEC60601-1:2012
MDD Classification
Class IIa (continuous operation mode)
EMD compliance
Complies with EMD standard IEC60601-1-2:2014.
Type of protection
NIBP: type B
Applied part
Type of protection against
electrical shock
Class I
Blood pressure measurement
Measurement method
Oscillometric measurement
Pressure display range
0-299 mmHg
Pressure display accuracy
Pressure: ±3 mmHg
NIBP Measurement range
SYS
DIA
Pulse rate
40-270 mmHg
20-200 mmHg
30-240 bpm
NIBP Clinical test
EN1060-4
:2004
Pulse rate accuracy
±5%
Cuff
Winding mechanism operated by geared motor
Applicable arm circumference 18-35 cm
Inflation
Automatic inflation by air pump
Deflation
Automatic deflation by mechanical exhaust
Rapid deflation
Automatic rapid deflation by solenoid valve
Environment specifications
Operating environment
Temperature: 10-40 °C
Humidity: 15-85% RH (no condensation)
Storage environment
Temperature: -20 to 60 °C
Humidity: 10-95% RH (no condensation)
Atmospheric pressure range 70-106 kPa (both for operation and storage)
Physical specifications
External dimensions
241 (W) x 324 (H) x 390 (D) mm
Weight
Approx. 5.5 kg
Functional specifications
Display method
3-digit display LED & LED lamp
Printer
Thermal printing, paper width: 58 mm
Usable life
5 years from installation
According to A&D data (tested for use under recommended
environment, including maintenance inspection. Results may
be different under other conditions.)