13.2. Factory Settings
This function can set the following parameters to factory settings.
– All measurement programs
– All results stored in memory function.
– All parameters of the function table
– ID number is reset to 0000000.
– Order of calendar and date.
13.2.1. Operation
1. Turn on the analyser.
The display will show 0.000 g.
2. Press and hold the SELECT key to enter the function table.
3. Press the SELECT key several times to display
CLr
.
4. Press the ENTER key to enter the item.
5. Press the
or
key to display
CLr Go
.
Caution
Press the ENTER key whilst
CLr no
is displayed and the
operation will be cancelled.
6. Press the ENTER key to reset.
End
is displayed.
7. Press the RESET key to return to the weighing mode.
0.000
CL AdJ
CLr
SELECT
SELECT
ENTER
ENTER
RESET
0.000
CLr no
CLr Go
End
CL AdJ