MS-70 / MX-50 / MF-50 / ML-50
33
7.2.2. Ramp
Drying
Standard Mode, Automatic Mode or Manual Mode
Refer to page 35 for Timer Mode
This explanation uses the following parameters and sample displays of MX-50.
Common
Items
Program
number ..........................................3 ( PROG 3 )
Drying
program
Heating pattern....................Ramp drying (
)
Final drying temperature.....160°C
Ramp time...........................5.0 minutes
Measurement
Unit..............................................Moisture content
Minimum scale value during measurement ..0.01 %
Minimum scale value of gram display...........0.001 g
Sample
quantity............................................Approximately 5 g
Data
memory
function ..................................Not used
Items for Standard Mode
Analysis
mode ..............................................Standard mode (Symbol:
5td
)
ACCURACY .................................................MID.
Analyzing mode to finish measurement............Automatic setting by ACCURACY
Items
for
Automatic
Mode
Analysis
mode ..............................................Automatic mode (Symbol:
U-a
)
Analyzing mode to finish measurement............0.05 %/min
Items for Manual Mode
Analysis mode ..............................................Manual mode (Symbol:
U-m
)
1. Display gram unit (of the weighing mode).
Select a Program Number to Edit the Measurement Program
2. Press
the
PROGRAM
key and press the
↓
or
↑
key to select a program number.
3. Press
the
ENTER
key to use the number.
4. The analyzer displays
end
and returns to the
weighing mode.
Caution If the data memory function is active, the data
memory number (MEM) is displayed in place of the
program number (PROG).
PROG
1
0.000
g
PROG
"3
end
PROG
3
0.000
g
Program number
% MOIST
D
/W
/D
Ramp time
Final drying temperature
(ML-50 can not select ramp drying)