38
Calibration test report using an external weight
Note
Calibration test does not perform calibration.
Key operation
1 Press and hold the
CAL
key to display
CC
out
and release the key.
2
Cal
0
is displayed.
3 When updating the target value, press the
RANGE
key and proceed to step 4.
When using preset target value, proceed to step 5.
4 Specify calibration mass value using the following keys.
RANGE
key ........ The key to select the blinking figure
RE-ZERO
(+)key .. The key to increase the value of the blinking figure.
MODE
(-)key........ The key to decrease the value of the blinking figure.
key ......... The key to store the new value.
5 When pressing the
RE-ZERO
key, the zero point is measured and the weight value is
displayed for a few seconds.
6 Place the displayed mass on the pan. Press the
key to measure the mass and the
weight value is displayed for a few seconds.
7 If GLP output is used,
glp
is displayed and the calibration test report is output.
8 The balance returns to the weighing mode after this test.
Setting of "
info 1
"
Setting of "
info 2
"
AD-8121 printer format
General format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR
Carriage return, ASCII 0Dh
LF
Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
-----------
A
-
&
-
D
MODEL
----
HR-300i
S/N
-----
01234567
ID
------
LAB-0123
DATE
--
2004/09/30
TIME
----
12:34:56
CAL.TEST(EXT.)
ACTUAL
-------
0.0000
--
g
----
+200.0002
--
g
TARGET
----
+200.0000
--
g
SIGNATURE
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
The space
to fill date in.
The space
to fill time in.