22
8. FUNCTION SWITCH AND INITIALIZATION
8-1 Permit or Inhibit
The balance stores parameters that must not be changed unintentionally. There are two switches for
the purpose of protecting these parameters. Each switch can select either “permit” or “inhibit”. The
“inhibit” protects parameters against unintentional operations.
FZ-CT/
FZ-GD
series
FX-CT/ FX-GD series
Function
switch
- Function table
- Calibration using the internal mass
- Calibration using an external weight
- Internal mass correction
- Function table
- Calibration
Switches (FZ-CT/ FZ-GD
series)
Switches (FX-CT/ FX-GD series)
Function table
0
To inhibit changes to the function table
1
To permit changes to the function table
Calibration using the internal mass
(One-touch calibration)
0
To inhibit calibration using the internal mass
1
To permit calibration using the internal mass
Calibration using an external weight
0
To inhibit calibration using an external weight
1
To permit calibration using an external weight
Not used
Internal mass correction
0
To inhibit correction
1
To permit correction
(The display shown left indicates the factory settings.)