69
19. SPECIFICATIONS
FZ-120
i
WP FZ-200
i
WP FZ-300
i
WP FZ-1200
i
WP FZ-2000
i
WP FZ-3000
i
WP
Weighing capacity
122
g 220
g 320
g 1220
g 2200
g 3200
g
Maximum display
122.084 g 220.084 g 320.084 g
1220.84 g
2200.84 g
3200.84 g
Minimum weighing value (1 digit)
0.001 g
0.01 g
Repeatability (Standard deviation)
0.001 g
0.01 g
Linearity
±
0.002 g
±
0.02 g
Stabilization time (typical at FAST)
Approx. 1 second
Sensitivity drift
(10°C to 30°C / 50
°
F to 86
°
F)
±2 ppm/°C
±
0.010 g
±
0.10 g
±
0.15 g
Accuracy right after calibration using the
internal mass. See notes below
*
The value above is to the weighing capacity.
Clock function
Buit-in clock
Operating environment
5°C to 40°C (41°F to 104°F), 85%RH or less (No condensation)
Display refresh rate
5 times/second, 10 times/second or 20 times/second
Minimum unit mass
0.001 g
0.01 g
Counting
mode
Number of samples
5, 10, 25, 50 or 100 pieces
Minimum 100% reference mass
0.100 g
1.00 g
Percent
mode
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
External calibration weight
100 g
50 g
200 g
100 g
300 g
200 g
100 g
1000 g
500 g
2000 g
1000 g
3000 g
2000 g
1000 g
Weighing pan diameter
130 mm
150 mm
Net weight
Approx. 3.2 kg
External dimensions
193 (W) x 262.5 (D) x 84.5 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power
receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Interface
RS-232C
Dust and water protection
Complying with IP65
*
Accuracy right after calibration using the accessory small breeze break and the internal mass,
in good ambient conditions (within the temperature range of 10
°
C to 30
°
C (50
°
F to 86
°
F) with
no abrupt changes in temperature or humidity, no drafts, no vibration and no effect by
magnetic fields or static electricity).
With the FZ-
i
WP series balance, the environment and passing age may cause internal mass
value error. Calibrate using an external mass and perform maintenance.