background image

 

 

 

FX

i

-

08 / GXM-08 / GXL-08

 

FX

i

-

08 / GXM-08 / GXL-08 

Ethernet Interface

 

 
1WMPD4001351E 

When this Ethernet interface (Below, this product) is installed into the A&D balances, the balances can communicate 
with a computer through the LAN port and the weighing data can be managed with the software "

WinCT-Plus

".   

Target models 

FX

i

-08

 

FX / FZ

FX-

 i 

/ FZ-

i

HR-A / HR-AZ

GX-A / GF-A Series 

GXM-08 

GX-M / GF-M Series 

GXL-08 

GX-L / GF-L Series 

 

Unpacking the Ethernet Board 

Ethernet Interface board 

 

 

 

 

 

 

 
 
 
 
 

 

Installing the Ethernet Board 

When setting the IP address, you may need the MAC address

*

  (hardware address) to identify the product 

individually, so make a note of the stated MAC address. 

 

*

 Example: 00

-

20

-

4A

-

84

-

23

-

6A 

Caution: 

Remove the AC adapter from the balance before doing the following work. 

 

  In case of 

FX

i

-08

   

Step 1  Remove the two screws from the rear of the balance and 

remove the panel. Peel the connector off that is attached on 
the panel with double-side adhesive tape. 

 

Step 2  Connect the connector to the this product. 

 

Step 3  Insert this product into the balance and screw the panel. 

 

  In case of 

GXM-08

 / 

GXL-08

   

Step 1  Remove the two screws from the panel and remove the 

RS-232C / USB board from the balance. 

 

Step 2  Pull gently and remove the two cables from the RS-232C / USB board. 

 

Step 3  Insert the two connectors into the product. 
 

Step 4  Insert the rubber packing and plug the this product into the balance. 

Screw the panel of this product. 
Caution 

  When inserting the this product, pay attention to the top and bottom orientation of the board. 

  The rubber packing has a deep groove on the balance side. 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 

Connecting the Ethernet Board to a Computer 

Caution 

  Contact the network manager before connecting the 

this product

 to a network.   

A network error may occur due to the 

this product

. A&D and dealers assume no responsibility for errors. 

  Preset the 

IP

 address and subnet mask into the 

this product

 one by one. The 

IP

 address, 172.16.100.2, is set 

for all 

this product

 at the factory. Do not duplicate 

IP

 addresses.   

  The 

IP

 address cannot be restored to factory setting.   

We recommend that you write the 

IP

 address on the supplied label. 

Preparation 

Please download each software from A&D website   

https://www.aandd.jp/products/software/software.html

for below. 

(1) 

Setting procedure “WinCT-Plus” instruction manual   

(2) 

IP address setting software “Device Installer” 

(3) 

Data acquisition software “WinCT-Plus” 

Step 1  Connect the Ethernet board and the computer using a hub and straight 

cables. (When connecting directly  without using  a  hub, use a cross 
cable)   

Step 2  Reconnect the AC adapter to the balance.

 

Step 3  Preset the IP  addresses and subnet mask of the this product  and the 

computer. For the setting procedure, refer to the “WinCT-Plus” instruction 
manual. 

 

A communication error may occur with the wrong operation. 

Step 4  Write the 

IP

 address on "

IP

 

address label" and affix it where you can 

see the address easily. 

Step 5  Install the data acquisition software to  refer to the “WinCT-Plus” 

instruction manual.   

IP

 address label (5 labels)

 

Write the IP address on a label and use it for identification. 

IP 

address protective cover (5 covers) 

Affix a protective cover over the IP address label.

 

AC adapter 

Step 2 

Hub

 

Step 1 

Computer 

Step 1 

Straight cable 

Cross cable 

Rear of the 

balance 

GXL-08 

Rubber 

packing 

 

 

Connector 

 

Connector with lock 

Screw 

RS-232C/USB board 

Connector 

MAC address 

 

Ethernet board 

 

Step 1 

 

Step 2 

 

Step 3 

Step 4 

Back side 

Step 2 

GXM-08 

Step 2 

Step 1 

 

Cable 

Connector with lock 

Screw 

Rubber 

packing 

Connector 

RS-232C/USB board 

Connector 

 

Step 3 

MAC address 

Step 4 

Ethernet board 

 

FX

i

-08

 

GXM

-

08

 

GXL-08 

MAC address 

MAC address 

MAC address 

( Connection example with 

FX

i

-08

 ) 

Cable 

Screws 

FX

i

-08

 

Panel 

Step 1 

Connector 

Step 2 

 

MAC address 

Ethernet board 

Step 3 

 

Reviews: