9
EN
6.3. Error code table
When any of the following errors are displayed, try turning the display off and on again.
Display
Content
error0
is displayed The temperature sensor has failed.
error3
is displayed
The memory (circuit) has failed.
error4
is displayed The internal circuitry has failed.
error5
is displayed The mass sensor has failed.
Summary of Contents for FG-150KCLWP