33
With nothing
on the pan
2.
CC 0
is displayed.
3. If it is necessary to change the target weight value that is used for calibration test,
press the SAMPLE key and change the value using the following keys.
SAMPLE key
To select the digit to change the value (blinking).
TARE key
To change the value of the blinking digit.
PRINT key
To store the value and return to step 2.
4. In step 2, press the PRINT key. The zero point
is weighed and the weight value is displayed for a
few seconds.
5. Place the target weight with the same value as
displayed on the pan. Press the PRINT key to
weigh it. The weight value is displayed for a few
seconds.
6.
end
appears.
7.
glp
is displayed and calibration test report is
output.
8.
end
appears again. Remove the weight and
press the MODE key to return to the weighing
mode.
CC 0
0.0
CC 0
The weighed data
is shown with “g”.
2000.0
2000.0
2000.0
Target
weight
end
end
The weighed data
is shown with “g”.
Returns to
the weighing mode
glp
GLP output