27
Calibration test report
The calibration test mode is used to compare a calibration test weight with the
calibration test data weighed by the balance.
This test does not perform calibration and this mode is available even after sealing
the calibration (CAL) switch.
1. Press and hold the SAMPLE and PRINT
keys.
CC
will appear. Then, release both
keys.
Pressing and holding the calibration (CAL) switch
will also display
CC
after
Cal
. Then,
release the CAL switch while showing
CC
.
2.
CC 0
is displayed.
3. If necessary, change the value of calibration test
weight according to the procedure step 4 of “7-1.
Calibration using a weight”.
4. With nothing on the pan, press the PRINT key.
The zero point is measured and the weighed value
is displayed for a few seconds. Then, the display
shows the value of calibration test weight.
5. Place a weight of the same value as displayed on
the pan and press the PRINT key to weigh it.
The weighed value is displayed for a few seconds.
6.
end
appears.
7.
glp
is displayed and calibration test report is
output.
8.
end
appears again. Remove the weight and
press the MODE key to return to the weighing
mode.
CC
CC 0
Press and hold both keys.
Release both keys.
0.0
1000.0
1000.1
end
glp
Returns to
the weighing mode
end
To output calibration
test report
To weigh