15
5. At step 3, pressing the PRINT key weighs the
zero-point value. Do not touch the pan during
weighing.
When the zero calibration is completed, the display
shows the calibration weight value.
To perform the zero calibration only and finish the
procedure, press the MODE key.
6. Place the calibration weight with the same value as
displayed on the pan. Press the PRINT key to
weigh it. Do not touch the pan during weighing.
7.
end
appears.
Remove the weight from the pan, and press the CAL
switch or MODE key to return to the weighing
mode.
Note
The value set in step 4 is stored in memory even after the power is
switched off.
If the balance is to be moved to other places, set the gravity
acceleration value of the area where the calibration using a weight is
to be done, and calibrate the balance according to the procedure
above. See the next section to set the value.
1000.0
end
Weighing pan
With nothing
on the pan
To store
To store
Calibration
weight
Cal 0
1000.0
Summary of Contents for EK-120i