39
AD-8127/AD-8121B
format
“
info 1
”
General format “
info 2
”
~
: Space, ASCII 20h
CR
: Carriage return, ASCII 0Dh
LF
: Line feed, ASCII 04Ah
Outputting the “Title block” and “End block”
When weight values are recorded as the GLP report, the “Title block” and “End block”
are added at the beginning and at the end of a group of weight values.
“Title block”
1. Press and hold the PRINT key.
Release the PRINT key when
5tart
is displayed. The
balance outputs the “Title block”.
2. Press the PRINT key or select
the auto-print mode to output the
weight values.
___________
A
_
&
_
D
MODEL
_____
EJ-123
S/N
____
Q12345678
ID
________
ABCDEF
DATE
__
2013/11/14
TIME
____
02:53:21
CAL.TEST(EXT.)
ACTUAL
________
0.000
__
g
_____
+100.001
__
g
TARGET
_____
+100.000
__
g
SIGNATURE
-
_
-
_
-
_
-
_
-
_
-
_
-
_
-
_
~~~~~~~~~~~
A
~
&
~
D
<CRLF>
MODEL
~~~~~
EJ-123
<CRLF>
S/N
~~~~
Q12345678
<CRLF>
ID
~~~~~~~~
ABCDEF
<CRLF>
DATE
<CRLF>
<CRLF>
TIME
<CRLF>
<CRLF>
CAL.TEST(EXT.)
<CRLF>
ACTUAL
<CRLF>
~~~~~~~~
0.000
~~
g
<CRLF>
~~~~~
+100.001
~~
g
<CRLF>
TARGET
<CRLF>
~~~~~
+100.000
~~
g
<CRLF>
SIGNATURE
<CRLF>
<CRLF>
<CRLF>
----------------
<CRLF>
<CRLF>
<CRLF>
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero value
Target weight value
Target weight
Signature
•
•
•
•
5tart
0.000
To output “Title block”
12.345
23.456
To output weight values
Press and hold
end
Summary of Contents for EJ-1202
Page 50: ...48 EJ 54D2 190 209 133 Min 114 Unit mm 50 ...
Page 53: ...51 World Map ...
Page 54: ...52 MEMO ...
Page 55: ...53 MEMO ...