![A&D EJ-1202 Instruction Manual Download Page 35](http://html1.mh-extra.com/html/aandd/ej-1202/ej-1202_instruction-manual_446376035.webp)
33
9.4.
EJ-13 Density Determination Kit
Using this option and calculation program,
the balance can determine the density
(specific gravity) of a sample.
EJ-13 is for use with the EJ-123 / EJ-303.
EJ-13 Installation
Remove the weighing pan and the fixing
frame.
Install the beaker stand on the balance
and attach the pan stand to the pan
support of the balance.
Remove the breeze break that can not be
used while using the density
determination kit.
Place a beaker filled with water on the
beaker stand and place the sample
weighing pan on top of the pan stand.
Density (Specific Gravity) Measurement
The density for a liquid can be set two ways.
One is to set the water temperature and the
other is to set the density value directly.
The factory setting for density of a liquid is
25 °C as water temperature (the density
value,
ρ
= 0.99704 (g/cm
3
), is used for
calculation).
The result is shown with three decimal places.
The density (specific gravity) is calculated by
the following formula.
S =
x
ρ
S: Density (specific gravity) of a sample
A: Weight in air
B: Weight in liquid
ρ
: Density of liquid (water)
Change the Function Table
Selecting a way to set the density of a liquid
Select the liquid density input method from the function table below.
The function table is available only when the density measurement mode is selected.
Class Item Param-
eter
Description
func ldin
Liquid density input
0
Water temperature
The way to input liquid density.
1
Liquid density
: Factory setting
A
A
-
B
Upper pan to weigh in air
Pan stand
Lower pan to weigh in water
Beaker stand
Pan support
Sample weighing pan
Weighing pan
Fixing frame
L
:
To weigh B in water.
A
:
To weigh A in air.
Breeze break