AD-4329A Weighing Indicator
Page
9
4.1.2. Rear Panel Description
No. Name
Description
15 Power
connector
Use accessory power cord.
Please Confirm that the local voltage and receptacle type are
correct for your scale.
16 Grounding
terminal
17 RS-232C interface
Refer to "12. RS-232C Interface" for the detail.
18 Blank panel
Option space ( for relay output).
19 Fuse
holder
Use 0.5A time lag fuse for AC100V to AC120V.
Use 0.315A time lag fuse for AC200V to AC240V.
20 External
input
Seven control inputs selected function by the F functions.
Refer to F10 to 16.
21 Serial
output
Printer port. Refer to "11. Standard Serial Data Output" for
the details.
22 Load cell connector Connect accessory load cell plug.
4.1.3. Other Displays and Symbols
Standby display.
Zero error when turning display on. If the
MODE
key is
pressed, a current weighing value may be displayed.
Over load display. Remove any load from the load cell
immediately. It may cause damage to the indicator.
Example of an error display.
4. Description of Panels and symbols
15 16
17
18
19 20
21
22