16
4.1. Changing the Weighing Speed Using WinCT-AD4212C
The weighing speed can be selected from the following three rates to minimize the influence on
weighing that is caused by drafts and/or vibration at the place where the balance is installed.
Indicator Speed Stability
FAST
MID.
SLOW
Fast Sensitive value
Slow Stable value
Operation
1. Press the MODE button.
2. Press the MODE button to select a weighing speed.
Either FAST, MID. or SLOW can be selected.
4.2. Calibration
Calibration
Calibration using the calibration weight.
Caution
Do not allow vibration or drafts to affect the balance during calibration.
Caution on using an external calibration weight
The accuracy of the weight can influence the accuracy of weighing. Select an appropriate
weight as listed below. A calibration weight (conforming to OIML, Class E2 or equivalent) is
provided with the balance as a standard accessory.
Model
Usable calibration weight
Calibration weight provided
AD-4212C-300 50g,
100g,
200 g
,
300g
AD-4212C-600 50g,
100g,
200 g
,
300g, 400 g, 500 g, 600 g
AD-4212C-3000
50g, 100g,
200 g
, 300g, 400 g, 500 g, 1000 g,
2000g, 3000g
AD-4212C-6000
200 g
, 500 g, 1000 g, 2000g, 3000g, 4000 g,
5000 g, 6000 g
AD-4212C-301 50g,
100g,
200 g
,
300g
AD-4212C-3100
50g, 100g,
200 g
, 300g, 400 g, 500 g, 1000 g,
2000g, 3000g
200 g
The calibration weight in bold type: factory setting
Display
This indicator means “the balance is measuring calibration
data”. Do not allow vibration or drafts to affect the balance
while this indicator is displayed.
Each pressing
switches
the indicators
MODE
MODE
Weighing speed setting