  Errors due to air flow and measures to take 

 Where the influence of ambient air flow is 

great such as: close to an air conditioner, door 
or passage way. Even very subtle air flow that 
is hard to be detected may influence the 
weighing operation.   

  Avoid those areas as a weighing site. 
  If weighing is to be performed in such an 

area, use a breeze break or take other 
appropriate measures. 

 

  Where the influence of heat or drafts is great 

 

Eliminate temperature differences 
between a sample and the environment.   
When a sample is warmer (cooler) than 
the ambient temperature, the sample will 
be lighter (heavier) than the true weight. 
This error is due to a rising (falling) draft 
around the sample.   

  Do not touch the sample directly with your 

hand. Use tweezers or other tools. 

  If you touch the sample, the same type 

error described above will occur. 

  Do not perform weighing where it is exposed to direct sunlight. Weighing errors may 

occur due to sudden temperature change or drafts.   

 

 

  Errors due to vibration and measures to take 

  Where the influence of vibration is great, such as:   

(1) Soft ground (2) Second or higher floor (3) Near center of a floor far from pillars (4) Seismic 
isolated structures (5) Near tall buildings. 
In the areas listed above, the scale may yield unstable weight values on windy days or after an 
earthquake. Especially in case of (4) and (5), weight values may be unstable for a long period of 
time because of long lasting low-frequency vibration after strong winds or an earthquake. 

 

 

  Errors due to other causes and measures to take 

  Change in temperature or humidity 

  A sudden change in temperature or humidity can generate a draft and cause the balance to 

absorb or exude moisture, which leads to unstable displays. 

  Avoid sudden change in temperature or humidity. 
  Use an air conditioner or humidifier to control the temperature or humidity. 

 

 

 Magnetic material 

The balance uses a strong magnet as part of the balance assembly, so use much care when 
weighing magnetic materials. 

  Place a non-magnetic object such as aluminum or brass between the sample and the 

balance to keep an appropriate distance between them while weighing. 

 
 

Draft

40 C

20 C

Summary of Contents for AD-4212C-300

Page 1: ...AD 4212C 300 AD 4212C 600 AD 4212C 3000 AD 4212C 6000 AD 4212C 301 AD 4212C 3100 Production Weighing Unit INSTRUCTION MANUAL 1WMPD4001992B...

Page 2: ...ranslated into any language in any form by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the instrument covered by this manual a...

Page 3: ...4 3 5 Power Supply 14 4 OPERATION OF WinCT AD4212C 15 4 1 Changing the Weighing Speed Using WinCT AD4212C 16 4 2 Calibration 16 4 3 Changing the Baud Rate Using WinCT AD4212C 18 5 SMART RANGE FUNCTION...

Page 4: ...utput the weighing digital data directly Therefore the AD 4212C series the weighing unit can be connected to a personal computer or a PLC directly z Dust protected and Protected Against Splashing Wate...

Page 5: ...sory RS 232C cable 10 m Personal computer USB converter Option AX USB 9P AD 4212C AD 4212C USB port Communication software WinCT AD4212C Accessory RS 232C cable 10 m AD 8922A Personal computer Communi...

Page 6: ...10 m Accessory RS 232C cable 10 m AD 4212C AD 4212C AD 4212C BCD cable Fabricate locally PLC AD 8922A 01 Connection cable PLC AD 8922A LAN cable PLC AD 8526 BCD output board Option AD 8922A 01 D sub...

Page 7: ...de reasonable protection against interference when the equipment is operated in a commercial environment If this unit is operated in a residential area it may cause some interference and under these c...

Page 8: ...RS 232C connector 1 Please confirm that the AD adapter type is correct for your local voltage and receptacle type 2 2 m and 5 m cables are available as options Refer to 10 OPTIONS 1 2 See note below 2...

Page 9: ...te screws with a size corresponding to an M5 screw The screw hole diameter 6 mm In case of securing the weighing unit from the under side use M6 screws with 1 mm pitch Notes The screws to secure the w...

Page 10: ...alance for 30 minutes or more with nothing on the weighing pan 6 Set the pan unit and I O unit to adapt to the peripheral system Set the following for the I O unit RS 232C Refer to 6 COMMUNICATION SPE...

Page 11: ...esin or magnetic materials when designing a weighing pan When a locally fabricated weighing pan is used the zero point may be shifted greatly right after the AC adapter is plugged in Perform re zero b...

Page 12: ...nstall the weighing unit where it is free of dust Install the weighing unit away from equipment which produces magnetic fields Install the weighing unit in a stable place avoiding vibration and shock...

Page 13: ...the support device and the object to be weighed will be the value specified in the table below Even a load with a moment of less than the specified value when applied to the balance may cause the ove...

Page 14: ...and that can be sealed and then weigh it When the sample is powdery When the balance is used in combination with a feeder for batch weighing of powdery samples samples may be charged by rubbing sampl...

Page 15: ...Errors due to vibration and measures to take Where the influence of vibration is great such as 1 Soft ground 2 Second or higher floor 3 Near center of a floor far from pillars 4 Seismic isolated stru...

Page 16: ...weighing pan and keep the conditions around the pan clean after weighing powdery fluid samples or metallic strips 3 4 After Use Avoid mechanical shock to the weighing unit Periodically calibrate the...

Page 17: ...RS 232C terminal 3 Set the COM port of the WinCT AD4212C on the personal computer and press the START button The personal computer will display the weight value Function of display symbols and button...

Page 18: ...weight can influence the accuracy of weighing Select an appropriate weight as listed below A calibration weight conforming to OIML Class E2 or equivalent is provided with the balance as a standard acc...

Page 19: ...that there is nothing on the pan and press the PRINT button The balance measures the zero point Do not allow vibration or drafts to affect the balance The balance displays the calibration weight valu...

Page 20: ...Baud Rate Using WinCT AD4212C Operation 1 Click the Settings tab 2 Select the baud rate to be used and click the Set button Factory setting 2400 Settings tab Set button 3 Completed appears when the s...

Page 21: ...on range standard range Model Range used Weighing range Available minimum weighing value Precision range 1 0 g to 51 g 0 0001 g AD 4212C 301 Standard range 51 g to 320 g 0 001 g Precision range 2 0 g...

Page 22: ...specification Transmission system EIA RS 232C Transmission form Asynchronous bi directional half duplex Transmission rate Approx 10 times second to 50 times second Data format Baud rate 2400bps to 192...

Page 23: ...municate by connecting TXD RXD and SG When connecting to a PLC such as the external equipment do not connect to the internally used signal line 3 Data output format This format consists of fifteen cha...

Page 24: ...ghing data continuously Starts the continuous output of the weighing data The C command stops the continuous output Commands to control the balance Command Description CAL Enters the calibration mode...

Page 25: ...ibration weight being weighed After the weight is removed the balance returns to the weighing mode Displays END S T 0 0 0 0 0 1 2 g R E 0 0 D P 0 0 0 C H C A L 0 C H S T 0 0 PRT command P CR LF R T S...

Page 26: ...2 g S T 0 0 0 0 0 0 1 g Personal computer Time Weighing unit R CR LF The weight value exceeds 51 g Place a sample S T 0 0 0 0 0 0 0 g S T 0 0 0 0 0 0 0 g S T 0 0 5 1 2 3 4 g S T 0 0 1 2 3 4 5 g Person...

Page 27: ...he local A D dealer if the balance needs service or repair Use the original packing material for transportation Do not use organic solvents to clean the balance Clean the balance with a lint free clot...

Page 28: ...d table and perform checking When the balance proper performance is confirmed refer to page 11 to set up the installation site Checking that the operating environment or weighing method is proper Oper...

Page 29: ...3 2 When Building into a System Is the sample of magnetic material such as iron Use much care when weighing magnetic materials Refer to 3 2 When Building into a System 8 2 Asking For Repair If the bal...

Page 30: ...2400 bps to 19200 bps WinCT AD4212C communication software provided Sensitivity drift 2 ppm C 10 C to 30 C 50 F to 86 F Operating environment 5 C to 40 C 41 F to 104 F 85 RH or less No condensation Al...

Page 31: ...software provided Sensitivity drift 2 ppm C 10 C to 30 C 50 F to 86 F Operating environment 5 C to 40 C 41 F to 104 F 85 RH or less No condensation Allowable eccentric load 0 3 Nm or less With the ce...

Page 32: ...3000 6000 301 3100 Unit mm 98 48 8 30 18 8 46 50 50 91 9 1 30 48 8 9 3 1 66 15 6 46 0 2 50 59 3 M 6 80 5 254 231 231 218 0 2 50 6 5 2 4 3 hol e 4 M 4 Size of when connected Approx 30 Pan support dimen...

Page 33: ...e AD 8922A can output the data using BCD AD 8922A 01 comparator AD 8922A 04 or analog out AD 8922A 06 by selection of the option installed in the AD 8922A AD 8923 BCD CC Remote controller The AD 8923...

Page 34: ...If those are found to be charged discharge them using the AD 1683 DC static eliminator AD 8121B Printer Compact thermal dot matrix printer Statistical function clock and calendar function interval pri...

Page 35: ...removed repetitively Usually expressed as a standard deviation e g Standard deviation 1 digit This means that the measured values fall within 1 digit at a probability of about 68 Stabilization time Ti...

Page 36: ...h rate 28 29 E Earth ground 12 Electrostatic field meter 32 Ethernet converter 2 31 F FAST 2 16 28 29 G Grounding terminal 6 I I O unit 28 29 Insulators 12 L LAN 4 Leveling foot 6 7 LF 20 21 22 Linear...

Page 37: ...2 SLOW 16 Smart range function 2 19 SMP command 19 22 Stabilization time 2 28 29 33 Stable value 16 33 Standard range 19 Standby indicator 15 Static charge 13 Static electricity 9 11 27 33 U U command...

Reviews: