 exposed conductive parts, however, extra pre- 

   caution should be taken to minimize the possibility   

  of 

shorting. 

 

Do not make measurement with the Battery Cover 

removed.

 

 

Disconnect the test leads from the instrument for 

current measurement. 

 
(1) Set the Function Switch to “

600A

” or “

1000A

” position. 

  (on DSA680, only “600A” is available) AC has been 

selected by default; press the SELECT key, when DC 

has been selected, to change it to AC. AC mark is 

displayed at the upper left on the display. 

 
(2) Press the trigger to open the transformer jaws and 

 clamp them onto the one conductor under test, then 

 

take the reading on the display. Pressing the 

 “

Hz/DUTY

” Key switches the indication in following 

 sequence. 

AC Current            Hz          DUTY 

 

 

Hz/DUTY

 Function requires 50A or more at AC600A 

Range and 350A or more at AC1000A range. 
 

CAUTION 

 

Max conductor size for DSA680 is approx dia. 

33mm and for DSA1020 is approx dia. 40mm.   

  During current measurement, keep the trans-

  former jaws fully closed. Other wise, accurate 

 measurements 

cannot 

be 

taken. 

 

5-2. DC Current Measurement

 

DANGER 

 

Never make measurement on a circuit in which 

voltage over DC600V exists to avoid getting 

electrical shock. 

 

Do not make measurement with the Battery Cover   

   removed. 

 
(1) Set the Function Switch to “

600A

” or “

1000A

” position. 

  AC has been selected by default; press the SELECT 

  key, when AC has been selected, to change it to DC. 

 

(only 600A is available on DSA680) 

  DC mark is displayed at the upper left on the display. 

(2) 

With the transformer jaws closed and without 

 clamping them onto the conductor, press the “

ZERO

” key 

to zero adjust the display. (

mark is displayed at   the 

upper right on the display.) 

(3) Press the trigger to open the transformer jaws and 

clamp them onto the one conductor under test, the 

conductor should be at the center of the jaws, then   take 

the reading on the display. 

(4) Set the Function Switch to an appropriate position 

 according to current under test. 

(5) Pressing  the  “

ZERO

” key again releases “

ZERO

” 

 function.  (

mark at the upper right on the display 

 disappears.) 

 
 
 

5-3. AC Voltage Measurement   

DANGER 

 

Never make measurement on a circuit in which 

voltage over AC600V exists to avoid getting 

electrical shock. 

 

Do not make measurement with the Battery Cover 

removed. 

 

Keep your fingers behind the barrier on the 

instrument during measurement. 

 

(1) Set the Function Switch to “

ACV

” position. 

(2) Connect the red test lead to V/

 terminal and the 

 black test lead to COM terminal. 
(3) Connect the test leads to the circuit under test. Take 

the reading on the display. Pressing the “

Hz/DUTY

” key 

while reading is indicated on the display switches the 

indication in following sequence. 

 

AC Voltage          Hz          DUTY 

 

 

CAUTION

 

 

 

Hz/DUTY Function requires AC40V or higher.   

 

To measure a frequency, measure the voltage on 
the electrical circuit in advance. Then press the Hz

/DUTY key to enter into frequency measurement. 

 

Readings of frequency may fluctuate or be 

influenced under noisy environment. 

 

5-4. DC Voltage Measurement 

DANGER 

 

Never make measurement on a circuit in which 

 voltage over DC600V exists to avoid getting 

electrical shock. 

 

Do not make measurement with the Battery Cover 

removed. 

 

Keep your fingers behind the barrier on the 

 instrument during measurement. 

 

(1) Set the Function Switch to “

DCV

” position. 

(2) Connect the red test lead to V/

 terminal and the    

black test lead to COM terminal. 
(3) Connect the red and black test leads to the positive 

(+)and negative (-) sides of the circuit under test 

respectively. Take the reading on the display. If the 

connection is reversed, the display indicates the “

-

“ mark. 

 

5-5. Resistance/ Diode/ Cont/ Capacity   

Measurement 

DANGER 

 

Never use the instrument on an energized circuit. 

 

Do not make measurement with the Battery Cover   

   removed. 

 

Resistance 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. 
(2) Connect the red test lead to V/

 terminal and the 

black test lead to COM terminal. Confirm “

OL

” is 

indicated on the display, and then short-circuit the tips of     

test leads to make the indication zero. 

(3) Connect the test leads to the both ends of the resistor 

under test. 

(4) Take the reading on the display. 

 

CAUTION 

 

Even if short the test lead tips, indicated value may 

not be zero. But this is because of the resistance of 

test leads and not a failure. 

 

When test leads are open, “OL” is indicated on the 

display. 

 

Continuity 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. “

“ has been selected by default; press the 

SELECT key to change it to “

Continuity

”  

Resistance        Diode          Cont            Capacity 

 
 

(2) Connect the red test lead to V/

 terminal and the 

 

black test lead to COM terminal. Confirm “

OL

” is 

 indicated on the display and short-circuit the tips of test   

leads.  

Indication should become zero and buzzer sounds. 
 
(3) Connect the test leads to the both ends of the 

conductor under test. The buzzer sounds, if the 

resistance under test is 100

 or less. 

 

Diode 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. “

 “ has been selected by default; press the 

SELECT key to change it to “

Diode

”  

Resistance     Diode 

     

Cont 

 

      Capacity 

 
 

(2) Connect the red test lead to V/

 terminal and the 

black test lead to COM terminal. 

 
 
 
 

(3) Connect the red and black test leads to the Anode and 

cathode of the diode under test respectively.   

Take the reading on the display. If the connection is 

reversed, the display indicates “

OL

”. 

 

CAUTION 

 

Some of diodes cannot be tested. Indication on 

 the display will be “OL”. 

  (Zener diode, LED and so on) 

 

Capacity 

(1) Set the Function Switch to “

/Diode/Cont/Capacity

” 

position. “

 “ has been selected by default; press the 

SELECT key to change it to “

Capacity

”  

Resistance    Diode 

   

Cont 

 

   Capacity 

 

 

(2) Connect the red test lead to V/

 terminal and the 

 black test lead to COM terminal. 

(3) 

Connect the test leads to the both ends of the 

capacitor under test. 

(4) Take the reading on the display. 

5-6 Temperature Measurement 

(1) Set the Function Switch to “

ºC/ºF

” position. 

(2) Connect the K-type Temperature Probe to the 

input terminal. Positive (+) side of Probe should be 

connected to V/

(3) 

Contact the Sensor (metal part) of K-type 

Temperature Probe to the object under test. Take the 

reading on the display. 

  Positive (+) side of Probe should be connected to V/

 

WARNING 

 

Never connect the Temperature Probe to an 

energized circuit. 

 

CAUTION 

 

Room temperature is indicated on the LCD when 

 setting the Function Switch to “

ºC/ºF

” position. In 

 

case that “

OL

” or anything other than room 

 temperature is indicated, something may wrong 

 with the instrument. Stop the use of instrument 

 immediately. 

 

There may be a break in Probe when indication 

 

isn’t changed if Sensor (metal part) of K-type 

 Temperature Probe is contacted with the object 

under test. 

 

6. Other functions 

6-1. Sleep Function 

(1) This is a function to prevent the instrument from being 

left powered on in order to conserve battery life. This 

function causes the instrument to enter Sleep mode 

about 15 minutes after the last key operation. To exit the 

Sleep mode, turn the Function switch to “

OFF

”, then to 

any other position. 

 

(2)  Sleep Function is disabled when;   
MIN/MAX or PEAK Function is selected. Continuous 

measurement is made with the Sleep Function being 

disabled. To activate Sleep Function again, disable the 

MIN/MAX or PEAK Function. 

CAUTION 

 

The instrument consumes small amount of battery 

power in the Sleep mode. Set the Function Switch 

to the OFF position after use. 

 

6-2. HOLD Key

 

(1) Data Hold Function 

This is a function to freeze the measured value on the 

display. Press the “HOLD” key to freeze the   reading.   
The reading will be held regardless of subsequent 

variation in input. “

H

” is indicated on the upper left corner 

of the display while the instrument is in the Data   Hold   

mode. To exit Data Hold mode, press the “HOLD” key 

again. 

 

 
 
 
 
 
 

Anode

Cathode 

CAUTION 

 

Held readings are released when Sleep Function 

is activated while the instrument is in the Data Hold 

mode. 

 

(2) Backlight ON/OFF 
Pressing the HOLD key 2 sec or more lights up the 

Backlight. Pressing the HOLD key 2 sec or more again 

turns off the Backlight. 

 

6-3. NCV Function 

Red LED on the upper area on the Panel lights up 

at all ranges except for OFF when electric field   

exceeding 100V is detected by the sensor installed in   

the Jaws. It indicates a presence of voltage in an   

electrical circuit or equipment without 
touching them.NCV Sensor 
can detect electrical field   
only from the direction 
indicated in the right 
figure.Put the fixed   
element (left side) 
closer to the conductor 
under test. Detection against   
in-wall outlet is impossible.

 

 

DANGER 

 

The LED may not light up due to installation 
 condition of electrical circuit or equipment. Never 
 

touch the circuit under test to avoid possible 

danger even if the LED for NCV doesn’t light up. 

 

Check the functionality of LED on a well-known 
 power supply prior to measurement. When the 
LED doesn’t light up, do not make measurement. 

 

NCV indication is affected by external voltage, how 

to hold or place the instrument. 

 

6-4. MIN/MAX Function   

 

CAUTION 

 

SELECT, ZERO, Hz/DUTY keys are disabled 

 

while MIN/MAX Function is being activated. 

 

(1) 

AC/DC Current Range

 (600A only on DSA680) 

Pressing the MIN/MAX Key at 600A & 1000A Function 

enables min or max value measurement. Press the 

MIN/MAX Key to select MAX or MIN. The max or min 

value within measuring range is being held until this 

function is disabled. “MIN” or “MAX” is indicated on the 
display while this function is being activated. To disable 

this function, press down the MIN/MAX Key at least 2 

sec or change functions.   

 

(2) AC/DC Voltage Range 

 

CAUTION 

Pressing the MIN/MAX Key without applying voltage 

disables the Auto-ranging function and fixes the 

Range to 6V. Connect the test leads to the circuit 

under test and press the MIN/MAX Key after an 

appropriate range is selected by Auto-ranging 

function. 

   

Pressing the MIN/MAX Key enables min or max value 

measurement. Press the MIN/MAX Key to select MAX 

or MIN. The max or min value within measuring range is 

being held until this function is disabled. “MIN” or “MAX” 

is indicated on the display while this function is being 

activated. 
 To disable this function, press down the MIN/MAX Key 

at least 2 sec or change functions. 

 

6-5. ZERO Function 

 

CAUTION 

MIN/MAX, PEAK keys are disabled while ZERO 

Function is being activated. 

 

Zero Adjustment Function at Current Range “

” mark is 

to be indicated at the upper right on the display while 

ZERO function is being operated.   
Indication of relative value at Current, Voltage, 

Resistance: 
Pressing the ZERO Key indicates REL (relative value) 

Press the ZERO Key to save the initial value at the start 

of measurement as a reference value. Then the 

difference between the later measured values and the 

reference value is indicated on the display. The 

Auto-ranging function is disabled, while this function is 

being activated, and the Range is fixed to the Range 

selected at the start of measurement. Relative value is 

indicated within following ranges. 

(Measuring range) = 
(Full-scale value at the fixed Range) – (initial value) 

 

    To disable this function, press down the MIN/MAX Key 
    at least 2 sec or change functions. 

 

6-6. PEAK Function 

(600A only on DSA680) 

(1) Set the Function Switch to “

AC Current

” Range and 

 clamp onto a conductor under test. 
(2) Pressing the PEAK Key indicates “

P

MAX

” on the 

 display and initiates measurement. 
(3) Readings indicates the PEAK of current crest value.   
 

When measuring sine wave, reading is about 

2

 

times of RMS value. 

 

 
 
 
 
 
 
 
 
 
 
 

(4) Press the PEAK Key at least 2 sec to reset the 

 indication or release PEAK Function. 
Buzzer sounds twice, and the Function is released. 

 

CAUTION 

 

PEAK indication for Crest value is up to 1500A.   

Error indication is given when exceeding this 

range value. 

 

Sleep Function is disabled when PEAK Function 

 is selected. Care should be taken when 

 performing 

continuous 

measurement.  

 

6-7. Over-flow indication

 

When the input exceeds the measuring range at each  

Function other than Voltage ,1000A and Temperature 

Range "

OL

" or "

-OL

" is indicated on the display. 

 

7. Battery Replacement 

 

WARNING 

 

To avoid electrical hazard, set the Function Switch 

 to  “

OFF

” and remove the test leads from the 

 instrument before trying to replace batteries.   

 

CAUTION 

 

Do not mix old and new batteries. 

 

Install batteries in correct polarity as indicated in the 

Battery Compartment. 

 

 
 
 
 
 
 
 

Replace the batteries when a Low Battery Voltage warning 

BATT

” mark is indicated on the display. Note that when the 

battery is completely exhausted, the display blanks without 

BATT

” mark shown.

 

(1) Set the Function Switch to “OFF” position. 
(2) Unscrew and remove the Battery Compartment cover 

on the bottom of the instrument. 
(3) Replace the batteries observing correct polarity. Use 

new R03 (AAA) or LR03 / 1.5V batteries. 
(4) Install the Battery Compartment and tighten the   screws. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

8. Maintenance

 

 

Cleaning

 

Use a cloth dipped in water or neutral detergent for 
cleaning the instrument. 
Do not use abrasives or solvents. Otherwise, instrument 
get damaged, deformed or discolored. 

 

 

 

Lifetime Limited Warranty 

The attention to detail of this fine snap-around 
instrument is further enhanced by the application of A.W. 
Sperry's unmatched service and concern for detail and 
reliability. 
These A.W. Sperry snap-arounds are internationally   
accepted by craftsmen and servicemen for their   
unmatched performance.   
All A.W. Sperry's snap-around instruments are 
unconditionally warranted against defects in material 
and workmanship under normal conditions of use and 
service; our obligation under this warranty being limited 
to repairing or replacing free of charge, at A.W. Sperry 
snap-around instrument that malfunctions under normal 
operating conditions at rated use.*1 

 
Replacement procedure 

Securely wrap the instrument and its accessories in a 
box or mailing bag and ship prepaid to the address   
below.  
Be sure to include your name and address, as well the 
name of the distributor, with a copy of your invoice from 
whom the unit was purchased, clearly identifying the 
model number and date of purchase.

 

 
A.W.SPERRY INSTRUMENTS INC. 
ATT: Customer service dept. 
2150 Joshua's Path, Suit 302, 
Hauppauge, N.Y. 11788 

*1 .The warranty is not applicable if the instrument 
has been: misused, abused, subjected to loads in 
excess of specifications, has had unauthorized 
repair or has been improperly assembled or used. 
*Note: Recommended calibration interval should not   
exceed one year. Calibration service charges are not   
covered terms and conditions of warranty. 
 
 