background image

GlucoMen

®

 LX  -  38973 - 09/07

UK

GlucoMen

®

 LX  -  38973 - 09/07

TeS�TinG a blOOd S�aMple

5.  Use the lancing device, loaded with a new lancet, to puncture the finger. (See lancing device 
 

instructions for use.)

6.  Gently squeeze your finger to help form a drop of blood.
7.  Touch the end of the test strip to the blood drop until the test strip is full and the on-screen 
 

countdown timer begins. (Beeper sounds if enabled.)

nOTe

The Blood Drop symbol flashes on and off repeatedly until sufficient blood has been added 
to the test strip.

8.  A countdown on screen appears while test is in progress. Glucose result is available 
 

on-screen in 4 seconds. 

9.  The result is automatically stored into memory.
10. Press the Left/Right 

 buttons to move between marked (

) or unmarked results. 

Marked results are not included  in the average test results. HI, LO, and Control results 
are not included in the average.

11. Press the Mode 

 button to save the Marking Status: Marked (

) or Unmarked.

12. If  test  result  is  above  33.3  mmol/L,  the  screen  displays  HI;  test  result  cannot 

be stored. If test result is below 1.1 mmol/L, the screen displays LO; test result cannot 
be stored.

TeS�TinG a blOOd S�aMple

nOTe

The meter will time out after 2 minutes of non-use or if the strip is removed. The keys are disabled until a 
strip is inserted. Results and marking status are saved if the meter times out, the strip is removed, or the 
meter is turned off.

nOTe

You must apply the sample to the GlucoMen

®

 LX Sensor test strip while the blood drop symbol is flashing 

in the display. If the meter turns off before sample application, remove the GlucoMen

®

 LX Sensor test strip 

and insert again.

nOTe

If blood is not applied within 2 minutes, the meter will go into “sleep mode” to conserve the battery life.

nOTe

Lancets are for one-time use only. Use a new, sterile lancet each time you test. Test different areas on 
your fingertips to avoid developing calluses. Remove the used lancet from the lancing device. Follow your 
local disposal regulations where applicable.

WarninG

Your lancing device is for your personal use only. DO NOT share with others. Sharing the lancing device or 
lancets can transmit very serious infections. To avoid accidental fingersticks injuries, do not store used lancets 
in the device after testing or arm the lancing device with a new sterile lancet unless ready to use.

Summary of Contents for GLUCOMEN LX

Page 1: ......

Page 2: ...orks Please read it carefully before us ing your new meter The GlucoMen LX blood glucose meter is designed to be conve nient and easy to use It gives accurate results in just 4 seconds using a very sm...

Page 3: ...American Diabetes Association ADA the US Food and Drug Administration FDA and the Advanced Medical Technology Association AdvMed Notes Cautions and Warnings NOTES Provide helpful operating informatio...

Page 4: ...itical to your care Since any meter may fail break or be misplaced you should always have a backup meter This product may become infectious in the course of use Discard the product in accordance with...

Page 5: ...tive 98 79 EC on in vitro diagnostic medical devices SN Serial number In Vitro Diagnostic Medical Device Direct Current Voltage In vitro device for self testing Catalogue number Blood glucose meter Bl...

Page 6: ...process provides a blood glucose result Test results are available in 4 seconds There is memory for one user s set of test results 400 The meter is powered by a battery that can perform approximately...

Page 7: ...0 non condensing INTRODUCTION Before Testing Before testing and to ensure accurate glucose results wash your hands and the testing site then thoroughly dry these areas The GlucoMen LX blood glucose me...

Page 8: ...frigerate or freeze Do not store near heat or moisture Store the test strips in their original vial only After removing a test strip from the vial immediately replace the vial cap and close tightly Do...

Page 9: ...ou have completed a blood glucose or control solution test the last result will appear the next time your meter is turned on The date and time displayed is the date and time of your previous blood sug...

Page 10: ...onsistent with how you feel To check the performance of the meter and test strips RUNNING CONTROL SOLUTION Important Information for Control Solution Use only the GlucoMen LX Control solution for the...

Page 11: ...solution vial Discard a drop before use Squeeze a drop of control solution onto a clean hard dry surface e g control cap 4 Pick up the meter with test strip inserted and touch the test strip to the co...

Page 12: ...range is ei ther higher or lower your meter or test strip may not be working properly Repeat the process using a new test strip Do not use the meter until test results fall within the appropriate rang...

Page 13: ...mol L the screen displays LO test result cannot be stored TESTING A BLOOD SAMPLE NOTE The meter will time out after 2 minutes of non use or if the strip is removed The keys are disabled until a strip...

Page 14: ...may cause the blood drop to smear 1b Select the Palm Heel or Palm Side test site in the highlighted areas as shown 2 Attach the clear cap to the lanc ing device Press and hold the lancing device FIRML...

Page 15: ...forearm and palm testing TESTING A BLOOD SAMPLE 4 Continue to hold the device pressing FIRMLY until a proper blood drop forms The clear cap allows you to see the blood drop 5 Apply blood drop to the...

Page 16: ...lthcare professional and or contact your healthcare professional immediately If you receive an Error Message see page 38 of this user manual displays meaning actions TESTING A BLOOD SAMPLE Do not use...

Page 17: ...s 2 Press the Left Right button to view all the data in memory The Left arrow goes back in time and the Right arrow goes forward in time All results including control results marked results and unmark...

Page 18: ...will turn off automatically after 30 seconds NOTE When the meter memory is full 400 test results each new test result stored in memory will remove the oldest test result stored in memory DATA MANAGEM...

Page 19: ...e the battery icon in the upper right corner of the screen and the blood drop at the lower left corner of the screen the meter has sufficient charge for 20 more tests Continue with testing as usual th...

Page 20: ...er is ready to ac cept blood Apply a blood sample to the test strip Refer to how to run a blood sample Countdown screen 4 second countdown as meter calculates the blood glucose test result No action r...

Page 21: ...TIONS Display What it Means What to Do Your blood glucose reading is higher than 33 3 mmol L You may have high blood sugar Retest your blood glucose immediately If your reading is still HI you should...

Page 22: ...the required testing temperature range of 5 to 45 C 41 to 113 F Move the meter to a warmer or cooler area as appropriate and wait a few minutes DISPLAYS MEANINGS ACTIONS Display What it Means What to...

Page 23: ...new strip Blood Sample Error Incorrectapplicationofbloodsampleorcontrol solution onto the test strip or the test strip may be damaged Review your sampling technique The blood sample quantity may be in...

Page 24: ...st plasma The anticoagulant sodium heparin may be used EDTA is not recommended for use with GlucoMen LX Sensor test strips Interferences from elevated levels of Acetaminophen Tolaza mide Uric Acid Bil...

Page 25: ...not return to Data review mode 3 Battery low icon is displayed in every mode except setup 4 Once battery level goes below the threshold that triggers the low battery warning it continues to give the w...

Page 26: ...ns 3 We have no knowledge of the performance of the GlucoMen LX blood glucose meter when used with test strips other than GlucoMen LX Sensor test strips Therefore we make no warranty as to the perform...

Page 27: ......

Reviews: