33
Per
for
ma
nce C
ha
ra
ct
er
is
tics
/
English
/ 10.6
β
-Ketone: Accuracy
The accuracy of the the GlucoMen
®
Day METER 2K system was
assessed by comparing blood
β
-ketone results obtained by patients
with those obtained using a Randox Monaco, a laboratory instrument.
Slope
0.9922
Y-intercept
0.03908 mmol/L
Correlation coe cient (r)
0.994
Number of subjects
200
Range tested
0.01-7.66 mmol/L
/ 10.7
β
-Ketone: Precision
The precision studies were performed in a laboratory using GlucoMen
®
Day METER 2K system.
Within Run Precision
*Bloodav 0.5 mmol/L
SD = 0.039 mmol/L
*Bloodav 1.3 mmol/L
SD = 0.061 mmol/L
*Bloodav 3.5 mmol/L
CV = 3.9 %
*Bloodav 5.6 mmol/L
CV = 3.7 %
*Bloodav 7.3 mmol/L
CV = 4.5 %
Between Run Precision
*Controlav 0.80 mmol/L
SD = 0.017 mmol/L
*Controlav 2.65 mmol/L
CV = 2.7 %
*Controlav 4.74 mmol/L
CV = 3.2 %
This study shows that there could be variation of up to 4.5 %.