PT300, Service Manual
Rev B, October, 2000
22
System checkout procedure
The following measurements are typical values. When tolerances are given the signal
must fall within the given ranges. Special measuring conditions will be specified as
needed. The measurements are given in the order that you should use them.
Condition:
Scale is off and plugged into the charger. No cells installed. A variable
power supply connected to the power supply leads, set to 12.0V.
Charger voltage:
Positive battery lead, 8.6V
To check out the charger circuit, install cells.
Positive battery lead, 5.4V to 8.6V, depending on batteries state of charge.
On / Off switch:
Press the ON switch,
Q6, collector, 6.8V or greater.
Press the OFF switch.
Q6, collector, 0.7V or less.
Press the ON switch for the following tests.
Power Supplies:
U19, Pin 8 = 5.0V ±0.25V
U19, Pin 5 = -4.9V ±0.25V
U10 pin 5, 0.5mS/Div
U2 pin26, 5mS/Div
U3 pin 4, 0.2mS/Div
Figure 9 Test Waveforms
All signal measured with vertical gain set to 1 volt/division. For the reset pulse, ground
and release the reset pin while watching with the oscilloscope.
CAUTION:
Summary of Contents for PT300
Page 1: ...I PT300 Service Manual A A Scales 1 813 994 3190 Fax 509 355 3498 www stainlessscales com...
Page 45: ...PT300 Service Manual Rev B October 2000 44 Appendix B Schematics...
Page 46: ...PT300 Service Manual Rev B October 2000 45...
Page 47: ...PT300 Service Manual Rev B October 2000 46...
Page 48: ...PT300 Service Manual Rev B October 2000 47 Totalizing cable...
Page 49: ...PT300 Service Manual Rev B October 2000 48 Charging Cable...
Page 50: ...PT300 Service Manual Rev B October 2000 49 Loadcell and Summing Board...
Page 51: ...PT300 Service Manual Rev B October 2000 50 Internal Cable Assembly...
Page 52: ...PT300 Service Manual Rev B October 2000 51 Parts...