![5Pascal LIOSMART 100 User Manual Download Page 18](http://html1.mh-extra.com/html/5pascal/liosmart-100/liosmart-100_user-manual_3067730018.webp)
User Manual
LIOSMART 100
18
Press
1
To choose between programs press on the number (below display) or use and press .
In the process
Meaning:
STEP Nr: 1: 10:00
First step in the process. Time set to 10 minutes [when temperature
reaches the tolerance area and parameter Tsh/Tm in in set limit the
time count down starts. See special settings]
Tsh=26.0°C / -20.0°C
Current shelve temperature is 26.0°C, set temp. -20.0°C
P= 1015.0mb / 1000.0mb Current chamber pressure is 1015.0 mbar set pressure is 1000 mbar
Tm = 26.8°C
Current temperature on material
Tc = 19.0°C
Current condenser temperature
*Vacuum value from Pirani vacuum gauge. All values above 10 mb are treated as atmospheric
pressure. Only values below 10 mbar can be considered as correct measured value!
WARNING:
This kind of display indicates automatic running of selected program.