4Deep Desktop Microscope User Manual Download Page 8

The distance between the camera sensor and the point source - Source to Screen Distance
(SSD) of the Desktop microscope is adjustable. Contact 4Deep for the development of custom
configurations for special applications.

In order to achieve accurate images, inline holography requires that a reasonable amount of
the reference wave (unscattered light) reaches the camera sensor. Too many objects between
the PS and the camera can reduce the amount of reference wave reaching the camera and will
therefore result in poor reconstructions.

For moving objects, keep the exposure time of the camera as short as possible while still
maintaining a visible image. Long exposure times will result in blurry or smeared images. The
exposure can be adjusted in the camera control of the Octopus software.

The PS does not produce a large amount of light; therefore background light from external
sources must be reduced in order to obtain optimal images. Excessive background light can
wash out the images and produce low contrast reconstructions.

If possible, always obtain a background image (a hologram containing only the light from the PS,
without a sample/object causing scatter). A good background image will allow for background
subtraction when reconstructing holograms, producing higher contrast reconstructions while
reducing noise.

9

Care and Maintenance

Your Desktop microscope requires very little maintenance. Keeping the microscope casing
clean and free of lint or dust is recommended. Storing the microscope protected from dirt, dust
and moisture will ensure a long operational life.

Take all measures to prevent contact with

the point source, as any dust or liquids on the pinhole may prevent the proper operation
of the microscope.

10

Troubleshooting

Problem: Camera is not detected by the software.

Solution: Make sure the power supply connector is properly connected to the microscope.
Make sure the microscope is connected to the Ethernet port on your computer, and that the
power button is on and indicating a current with the light.

Problem: No light from the point source.

Solution: If there is no light coming from the point source and it is properly connected there may
be an alignment problem with the laser, there may be direct damage to the pinhole, or the laser
diode may have burned out. Send the microscope or point source back to the manufacturer for
re-alignment.

8

Summary of Contents for Desktop Microscope

Page 1: ...Desktop Microscope User Guide Version 1 1 0 4Deep inwater imaging 6589 Chebucto Road Halifax Nova Scotia B3L 1L9 Canada tel 902 444 4449 info 4 deep com www 4 deep com 1...

Page 2: ...oscope and its Components 3 4 1 Camera 4 4 2 Sample Stage 5 4 3 Point Source 5 5 Microscope Setup 6 5 1 Step 1 6 5 2 Step 2 6 5 3 Step 3 6 6 Principle of Operation 6 7 Software 7 8 Tips for Optimal Ho...

Page 3: ...viceable parts inside the light source If problems are encountered regarding the light source please return it imme diately for service Allow only a qualified technician to disassemble the point sourc...

Page 4: ...rolled by the camera vertical control knob shown in Fig 2 This alters the Source to Screen Distance SSD which changes the magnification and resolution of the microscope Increase SSD Magnification Incr...

Page 5: ...ject Distance SOD and the vertical position of the sample stage controls this Warning Be careful of the point source when moving the vertical sample stage position DO NOT touch the point source a The...

Page 6: ...h the switch shown in Fig 4 on the previous page 5 3 Step 3 Open the Octopus program and go to the camera tab The program connects to the camera automatically and holograms can be saved or reconstruct...

Page 7: ...scope camera at the rate of up to 15 frames per second Octopus provides advanced 3D visualization and particle detection capabilities Stingray is an automated particle recognition and morphology class...

Page 8: ...containing only the light from the PS without a sample object causing scatter A good background image will allow for background subtraction when reconstructing holograms producing higher contrast rec...

Page 9: ...erating system Windows XP or later Communication power Ethhernet 12 VDC Hologram size from 2048x2048 to 512x512 Frame rate from 16 to 50 fps List of Figures 1 Main view of the Desktop microscope 4 2 T...

Reviews: