AS II Errors
The AS II will display error codes when there is a problem with the strip or proof-test.
Clamp Slipping
When the AS II tries to pull the fiber to the set tension value and that value cannot be
achieved, the unit will display ‘Clamp Slipping’.
Note:
When using rubber in the clamps, this will affect the maximum tension that can be
achieved during the proof-test. It may not be possible to achieve the 20N maximum
1 Press the
Go Button
to clear the error.
2 Press the
Clamp Buttons
to release the fiber.
Fiber Too Tight
This will only occur when proof-testing. If the fiber is loaded into the clamps and the fiber is too
tight, the AS II will beep once
(for the closing of the clamps)
, then beep 3 more times and
display ‘Release Fiber’. Release the fiber and try loading again.
Broken Fiber
When a fiber breaks during the proof-
test process, the ‘Broken Fiber’ error will be displayed.
1 Press the
Go Button
to clear the error.
2 Press the
Clamp Buttons
to release the fiber.