Figure 50 Calibration Validity
If power on calibration is selected, it could do white and black calibration every time when you turn on the
instrument. If not, it would be only able to check records and not complete the test.
If 4 Hours validity is selected, the validity of Calibration will expire in 4 hours. If Expired, it would be only able to
check records and not complete the test. Until re-calibrate it, the calibration is valid.
If 8 Hours validity is selected, the validity of Calibration will expire in 8 hours. If Expired, it would be only able to
check records and not complete the test. Until re-calibrate it, the calibration is valid.
If 24 Hours validity is selected, the validity of Calibration will expire in 24 hours. If Expired,it would be only able to
check records and not complete the test. Until re-calibrate it, the calibration is valid.
3.8.7 Control Mode
When the instrument communicates with the PC software, the user can set the specific measurement control
mode.
In the system settings interface, click "Control Mode" to choose as shown in Figure 51.
Key:
if selecting this mode, when the instrument is connected with PC software, it only could use keys to
complete the test and upload data to PC software.