NS810 SPECTROPHOTOMETER OPERATING MANUAL
10
4.3.3 Sample Measurement
After testing the standard, press “Sample Measure”, it will enter Sample
Measurement interface as shown in Figure 7. Put the measuring aperture on the
sample, press “Testing” key. The colorimeter will display color data of the
sample.
The L*a*b*C*h* data on the left side is current sample data. The data on the
right side is color difference data between current sample and standard.
Figure 7 Sample Measurement
Click “No Name”, it will enter “Edit Name” interface as shown in Figure 16. You
can edit the name of the current standard.
Click “D65”, it will enter “Light Source” interface as shown in Figure 19. You
can set the light source of the current standard.
Click “10” or “2”, you can switch the observer angle between 2° and 10°.
Click “LAB”, it will enter “Color Space” interface as shown in Figure 22. You
can change the color space of the current standard.