INSPECTING THE RESPIRATOR
16
Testing the Batteries
On respirators equipped with a distress alarm, the battery condition can be tested manually as follows:
1 Make sure the distress alarm is off (i.e., the cylinder valve is closed, and no sensor module lights are flashing).
2 Press and hold the manual reset (yellow) button on the console. Observe the final light color (green or red) in the
sequence to determine the battery status.
•
Green lights on the control console and sensor module indicate that sufficient battery power remains.
•
Red lights on the control console and sensor module indicate that the batteries are low and must be replaced
before the respirator is used again. 3M recommends replacing all batteries before the respirator is used again.
See “
Chapter 5: Replacing Batteries
TIP
The yellow light is a combination of the red, green, and white lights that appears yellow-orange or amber
from a distance. At close range, the individual lights may be visible.
NOTE
Do not use tools to close the cylinder valve. Over-tightening the cylinder valve may cause damage that
could result in air leaking from the cylinder.
TIP
For information about changing batteries in 3M Scott Fire & Safety components or accessories, see the user
instructions provided with the components or accessories.
P/N 595373-01
Rev
C
202004
Summary of Contents for SCOTT AIR-PAK X3 PRO
Page 2: ......
Page 8: ...CONTENTS vi P N 595373 01 Rev C 202004 ...
Page 20: ...INTRODUCTION 4 P N 595373 01 Rev C 202004 ...
Page 52: ...OPERATING THE RESPIRATOR 36 P N 595373 01 Rev C 202004 ...
Page 68: ...52 P N 595373 01 Rev C 202004 ...
Page 72: ...56 P N 595373 01 Rev C 202004 ...
Page 74: ...58 P N 595373 01 Rev C 202004 ...
Page 78: ...62 P N 595373 01 Rev C 202004 ...
Page 80: ...64 P N 595373 01 Rev C 202004 ...
Page 82: ...66 P N 595373 01 Rev C 202004 ...
Page 83: ......