(G) Turn the vacuum pump on. Press the end caps against the O-rings.
•
(H) While the end cap is pulled tight against the housing, place the V-Band
clamp on the ridge of the housing and the end cap and tighten the
clamp. Repeat this step for the other end.
NOTE:
Using ISO Ashby Cross oil on the nut will help get the nut
on and off more easily.
•
Turn off the vacuum. Remove the blind flanges from the shell connections
and proceed to tighten the center seal.
CENTER SEAL TIGHTENING
•
(I) Insert the center seal tool into the center nozzle. Turn the tool until you
feel the notch in the tool mesh with the nubs of the center nut.
•
Using a counter clockwise motion, slowly torque the center seal to
90 ft/lb. (122 Newton/meters). Repeat for other end.
PRESSURE TEST
•
(J) Bolt blind flanges on both lumen ports and one of the shell ports.
•
Bolt a flange equipped with a 0-100 psi gauge and hose connection.
•
Pressurize the housing to 60 psig with clean, oil free, air.
•
Isolate the pressurized housing and monitor the pressure to verify a leak
free seal. There should be no pressure decay over 30 minutes.
G
H
I
J
Compressed
Air Source
11