64220, Rev. 1/April 2019
3M™ GMI™ PS200-LMP AUTO BUMP/CALIBRATION STATION
PERFORMING A CALIBRATION/BUMP TEST
5-1
5. PERFORMING A CALIBRATION/BUMP TEST
5.1. Calibration
1. Verify two gas cylinders are connected to the ABC.
2. Verify that the ABC has completed its warm-up sequence and only one green LED is illuminated,
Figure 5-1: ABC ON Indicator
3. Open the front cover by depressing the cover latch (as illustrated in
).
Open the cover fully.
Figure 5-2: Open Front Cover
4. Verify that the PS200-LMP is OFF.
5. Locate the monitor’s inlet connector over the ABC’s gas supply nozzle and engage, as illustrated
Locate Monitor on Gas Supply Nozzle
NOTE: The ABC will accept the monitor with or without the blue rubber boot fitted.