12
3. Insert the banana plug on one end of the lead wire (f) into the ground jack on
the back of the Charge Analyzer 711.
4. Insert the other banana plug end of the lead wire (f) into the gold cup
electrode (a).
5. Ground the unit by attaching the ground wire with alligator clip on the rear
panel of the unit to an earth/electrical ground.
6. Press the on/off button to activate the Charge Analyzer 711. The display will
briefly indicate CHANGE MODE OF OPERATION then VOLTMETER IS
ENABLED (if cup has not been previously attached).
7. Press the “B” button (SET) to display OFFSET A.
8. Select OFFSET A by pressing the “A” button to display UPDATE OFFSET
menu.
Note: If you should press the “B” button (SETUP), you will have to switch off the Charge Analyzer 711
to get back into the Voltmeter mode.
9. Select YES by pressing the “A” button. GROUND CUP, READY >B
will appear on the display for verification that the cup electrode is indeed
grounded.
10. Select the “B” button to activate the automatic zeroing feature in the
Voltmeter mode. Display will indicate READ ALL OFFSET, UPDATING,
zeroing all ranges.
This completes the zeroing procedure. The display returns to the measurement mode.
Use
1. Unplug the lead (f) from the ground jack at the rear of the Charge Analyzer
711.
2. Plug the lead (f) into the cylinder electrode (c).
3. Select the desired voltage range by momentarily pressing the “A” button.
Note: If OVERFLOW occurs during the measurement select a higher range.
The Charge Analyzer 711 is now ready to monitor the effectiveness of static
control products, measuring electrical charge on people and conductive objects,
indicating in volts.