3M Breathe Easy Turbo Operational Manualline Download Page 3

 

 

Cartridge selection and attachment: 

Remove three filters/cartridges from their packaging and verify that they are the 
appropriate type for the application by examining the label attached to the 
filter/cartridge housing. 
 
For gross decontamination of patients use FR-57 cartridge # 453-03-02 from 3M. 
 
Remove screw caps from each filter/cartridge and retain them for future use.  
Ensure that the air inlet insert is seated inside each of the three rubber air inlets on 
the Turbo unit and that the threads and sealing surfaces are free of dirt and debris.  
Screw a filter/cartridge loosely into each of the three threaded inlets.  When all 
three filters/cartridges are in place, hand-tighten them so that an airtight seal is 
achieved between the neck of each filter/cartridge and the rubber molding inside 
the threaded inlet.  Dispose of used filters/cartridges in accordance with 
applicable regulations. 

Cleaning of equipment: 

1.

 

Surface clean the Turbo Unit at the end of the shift unit after each use by using a 
damp sponge soaked in an approved cleaning solution to wipe the exterior 
surfaces of the filter/motor-blower assembly, the breathing tube, the battery pack, 
and the support belt.  Allow them to air dry. 

2.

 

Surface clean the outside of the head cover with a disinfectant before storing at 
the end of the shift.  Allow to air dry. 

 
Preventive Maintenance:  
 
Battery recharging: The black and yellow (standard) batteries can remain on the charger 
indefinitely since this charger has a trickle charge feature  
 
Monthly PAPR inspections:  Inspect the 3m Breathe Easy respirators using the 3M 
Breathe Easy Turbo PAPR Assembly Monthly Inspection Checklist.  
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

Page 3 of 5; 3M Breathe Easy Turbo PAPR; 6/15/06 

Summary of Contents for Breathe Easy Turbo

Page 1: ...er until it snaps into place if the head cover is not already connected Reconnect the respirator headpiece breathing tube assembly to Turbo unit and verify that air is flowing before donning the respirator headpiece DONNING RESPIRATOR 1 Ensure that one end of breathing tube is attached to motor blower unit outlet tube and other end of the breathing tube is attached to the head cover 2 Adjust the b...

Page 2: ...unusable butyl hood and dispose of properly 9 Inspect the butyl hood for damage such as holes tears or broken seams If the hood is damaged in any way dispose of the hood 10 Clean disinfect and inspect the PAPR blower unit after each use Surface clean the blower unit by using a damp sponge soaked in an approved cleaning solution to wipe the exterior surfaces of the filter motor blower assembly the ...

Page 3: ...k of each filter cartridge and the rubber molding inside the threaded inlet Dispose of used filters cartridges in accordance with applicable regulations Cleaning of equipment 1 Surface clean the Turbo Unit at the end of the shift unit after each use by using a damp sponge soaked in an approved cleaning solution to wipe the exterior surfaces of the filter motor blower assembly the breathing tube th...

Page 4: ...er cartridge or incorrect filter cartridge selection d Examine the outside of the battery pack for cracks Replace if damaged 4 Breathing Tube a Carefully examine the entire breathing tube Look for tears holes or cracks b Bend the tube to verify that it is flexible 5 Airflow Check a With the breathing tube assembly disconnected from the Turbo PAPR blower unit and the system running insert the base ...

Page 5: ...Inspection and Record Keeping Breath Easy Turbo PAPR Date Inspected PAPR ID CE Turbo Unit Checked Breathing Tube Checked Airflow Rate cfm Comments Initials Page 5 of 5 3M Breathe Easy Turbo PAPR 6 15 06 ...

Reviews: