Standard Operational Guideline 

3M Breathe Easy Turbo Powered Air Purifying Respirator (PAPR)  

 

PURPOSE: 

To provide guideline instructions for usage of the 3M Breathe Easy Turbo 
Powered Air Purifying Respirator (PAPR) Respiratory System  
 
This PAPR system is used to provide respiratory protection for staff during 
various tasks such as the gross decontamination of chemically contaminated 
patients.  

 
EQUIPMENT LOCATION:  (REMAIN BLANK FOR EACH HOSPITAL TO FILL IN)
 
OPERATIONAL INSTRUCTION GUIDELINES: 
 
AIRFLOW RATE CHECK 

At the beginning of each shift perform flow rate check to ensure system is 
providing adequate air flow.  With the breathing tube assembly disconnected from 
Turbo blower unit and the system running insert the base of the flow meter into 
the Turbo unit outlet.  Be sure the Turbo unit and flow meter are resting in a 
vertical position.  Check that the center of the float rests at or above the 6 cfm 
mark.  If the flow rate is below 6 cfm, replace the cartridges and check flow 
again. 

Push the end of the breathing tube with the slotted hose connector into the rear of 
the butyl rubber head cover until it snaps into place if the head cover is not 
already connected. 

Reconnect the respirator headpiece/breathing tube assembly to Turbo unit and 
verify that air is flowing before donning the respirator headpiece. 

 
DONNING RESPIRATOR 

1.

 

Ensure that one end of breathing tube is attached to motor-blower unit outlet tube 
and other end of the breathing tube is attached to the head cover. 

2.

 

Adjust the belt length by pulling the whole belt to the right or left through the 
Turbo PAPR blower unit's bracket so the buckle centers at your waist (in front) 
and the Turbo PAPR blower unit is behind you. 

3.

 

To don the PAPR unit, place the blower unit back against your lower back (along 
your spinal column) with the breathing tube extending upward.  Fasten the belt 
around your waist at the front so the Turbo unit rests comfortably and securely 
against your lower back. 

4.

 

Plug the Turbo unit into a fully charged battery pack (see step 3 for appropriate 
battery to use) at the beginning of the shift and attach the battery pack to the belt. 

5.

 

Turn the battery pack on to draw ambient air through the filter/cartridge. 

6.

 

Pull the head cover over your head and adjust it so the headband wraps around 
your head and the elasticized edge of the face seal fits under your chin. 

Page 1 of 5; 3M Breathe Easy Turbo PAPR; 6/15/06 

Summary of Contents for Breathe Easy Turbo

Page 1: ...er until it snaps into place if the head cover is not already connected Reconnect the respirator headpiece breathing tube assembly to Turbo unit and verify that air is flowing before donning the respirator headpiece DONNING RESPIRATOR 1 Ensure that one end of breathing tube is attached to motor blower unit outlet tube and other end of the breathing tube is attached to the head cover 2 Adjust the b...

Page 2: ...unusable butyl hood and dispose of properly 9 Inspect the butyl hood for damage such as holes tears or broken seams If the hood is damaged in any way dispose of the hood 10 Clean disinfect and inspect the PAPR blower unit after each use Surface clean the blower unit by using a damp sponge soaked in an approved cleaning solution to wipe the exterior surfaces of the filter motor blower assembly the ...

Page 3: ...k of each filter cartridge and the rubber molding inside the threaded inlet Dispose of used filters cartridges in accordance with applicable regulations Cleaning of equipment 1 Surface clean the Turbo Unit at the end of the shift unit after each use by using a damp sponge soaked in an approved cleaning solution to wipe the exterior surfaces of the filter motor blower assembly the breathing tube th...

Page 4: ...er cartridge or incorrect filter cartridge selection d Examine the outside of the battery pack for cracks Replace if damaged 4 Breathing Tube a Carefully examine the entire breathing tube Look for tears holes or cracks b Bend the tube to verify that it is flexible 5 Airflow Check a With the breathing tube assembly disconnected from the Turbo PAPR blower unit and the system running insert the base ...

Page 5: ...Inspection and Record Keeping Breath Easy Turbo PAPR Date Inspected PAPR ID CE Turbo Unit Checked Breathing Tube Checked Airflow Rate cfm Comments Initials Page 5 of 5 3M Breathe Easy Turbo PAPR 6 15 06 ...